Angiotensin Converting Enzyme (ACE) Activity Assay Kit (MAES0334)

SKU MAES0334
Product Type Assay Kit
Size 100 Assays
Detection Method Colorimetric Method
Sensitivity 27.5 U/L
Range 27.5-682 U/L
$681

Request a formal quotation

A PDF quote with its own quote number, ready for your purchasing office to raise a PO against. Back within 4 working hours, from a PhD scientist — not an autoresponder.

  • PO-ready PDFQuote number, VAT, Incoterms
  • Volume pricing appliedDiscounts from 2 kits up
  • Valid 30 daysPrices held while you get sign-off
Anything we should know? optional

No account needed. We use your details only to prepare and send this quote. If it's urgent, say so in the box and we'll come back the same working day.

Global Shipping: 80+ Countries
White Glove Service: Available upon request
Batch Consistency: Contact Sales
Distributors: 60+ Countries

Description

system_update_altDatasheetsystem_update_altMSDS

Angiotensin Converting Enzyme (ACE) Activity Assay Kit

The Angiotensin Converting Enzyme (ACE) Activity Assay Kit (MAES0334) measures ACE1 activity in serum, plasma, and animal tissue homogenates by a colorimetric, kinetic spectrophotometric method. ACE cleaves the FAPGG substrate, decreasing absorbance at 340 nm, and the rate of decrease is used to calculate activity in U/L or U/g protein. The kit is a 50-minute, single-wavelength format intended for research use.

Angiotensin-converting enzyme 1 (ACE1) is the target measured by this kit. The enzyme acts on the substrate N-[3-(2-furyl)acryloyl]-L-phenylalanylglycylglycine (FAPGG), which is cleaved into FAP and GG. Activity is reported in units per liter (U/L) for serum and plasma and in units per gram of tissue protein (U/g prot) for tissue homogenates. One unit is defined as the amount of 1 µmol of substrate catalyzed per minute at 37°C.

The kit is listed under the research areas Amino Acids and Proteins and Metabolic Diseases, which defines the study contexts that the product supports. It is intended to measure ACE1 activity in serum, plasma, and animal tissue samples. The recommended dilution table covers human serum, mouse serum, rat plasma, and rat liver, kidney, and lung homogenates, which allows comparisons across these sample types. The product is for research use only and must not be used for clinical diagnosis.

The assay is a colorimetric method. FAPGG has its maximum absorption peak at 340 nm, and ACE1 converts it to FAP and GG, so absorbance at 340 nm decreases. Activity is derived indirectly from this decrease. In the procedure, 1000 µL of working solution is combined with 100 µL of sample or distilled water (blank) in a 2 mL EP tube, mixed, and incubated at 37°C for 1.5 min. The initial OD at 340 nm (A1) is read, the reaction is incubated at 37°C for 5 min, and the second OD at 340 nm (A2) is read in a 0.5 cm optical path cuvette. The change in absorbance is calculated as ΔA = A2 − A1.

Serum and plasma are tested directly. If testing does not occur on the same day, serum or plasma can be stored at -80°C for one month. Animal tissue is washed in cold PBS (0.01 M, pH 7.4) and homogenized at 20 mg per 180 µL of normal saline (0.9% NaCl) at 4°C. The homogenate is centrifuged at 10000 × g for 10 min, and the supernatant is kept on ice for detection. The protein concentration of the supernatant must be determined separately when activity is calculated per gram of protein. Reagents are equilibrated to room temperature before use, and the incubation temperature and time must be accurate.

Intra-assay precision was 3.0% CV and inter-assay precision was 4.4% CV, based on the product data. The analytical sensitivity is 27.5 U/L, determined by adding two standard deviations to the mean OD of the zero standard assayed 20 times. The detection range is 27.5 to 682 U/L. Recovery averaged 103% across three concentrations, with individual recoveries of 101%, 105%, and 103% at expected concentrations of 80, 230, and 600 U/L. In the appendix, intra-assay CV ranged from 2.8% to 3.2% and inter-assay CV ranged from 4.3% to 4.6% across three human serum samples.