Description
Glucose-6-Phosphate Dehydrogenase (G-6-PD) Activity Assay Kit
The Assay Genie Glucose-6-Phosphate Dehydrogenase (G-6-PD) Activity Assay Kit (MAES0064) measures G-6-PD activity in serum, plasma and animal tissue samples. The assay couples NADPH formation to WST-8 reduction and reads orange formazan at 450 nm on a microplate reader. It is intended for research use in enzyme and glycolysis-related studies.
Glucose-6-phosphate dehydrogenase (G-6-PD) is an enzyme that catalyzes the oxidation of glucose-6-phosphate to 6-phosphogluconate (6-PG). In this reaction, NADP+ is reduced to NADPH, so enzyme activity is linked to the production of this reducing cofactor. Assay Genie lists the kit under the synonyms G-6-PD and G6PDH, and the kit is intended to measure G-6-PD activity in serum, plasma, and animal tissue samples.
Researchers use this kit to quantify G-6-PD activity in studies of glycolysis and lipid metabolism, and in research on disulfidptosis, as listed in the product research areas. Because the assay covers serum and plasma from rat, rabbit, cynomolgus monkey, and human sources, and tissue homogenates from mouse and rat, it supports comparisons across species and tissues. Results are reported per liter for serum and plasma and per gram of tissue protein for homogenates, which allows tissue activity to be normalized to protein content.
The assay is colorimetric. G-6-PD reduces NADP+ to NADPH, and the electron coupling reagent 1-MPMS transfers electrons from NADPH to WST-8, which is reduced to an orange formazan with a maximum absorbance at about 450 nm. The formazan signal is proportional to G-6-PD activity in the sample. In the protocol, a sample working solution is prepared from substrate and chromogenic agent, 50 µL is added to each sample and standard well, the plate is mixed for 5 s, and it is incubated at 37 °C for 10 min before optical density is read at 450 nm.
Serum and plasma are tested directly. If they are not tested on the same day, they can be stored at -80 °C for up to one month. Tissue is homogenized at 4 °C in Reagent 1 (initial recommendation 20 mg per 180 µL), centrifuged at 10000 × g for 10 minutes at 4 °C to remove insoluble material, and the supernatant is kept on ice. Tissue results are normalized to protein, so the protein concentration of the supernatant must be determined separately. Kit components are stored at -20 °C for 12 months, and the chromogenic agent and standard powder are protected from light.
The product information reports a sensitivity of 0.01 U/L and a detection range of 0.01-50 U/L. The product page lists an intra-assay CV of 2.1% and an inter-assay CV of 5.8%. The manual reports intra-assay CVs of 1.9-2.4% and inter-assay CVs of 5.6-6.1% for three human serum samples with mean activities of 2.60, 11.50, and 38.50 U/L.
| Product Name: | Glucose-6-Phosphate Dehydrogenase (G-6-PD) Activity Assay Kit |
| SKU: | MAES0064 |
| Size: | 96 Assays |
| Sample Type: | Serum, plasma, animal tissue |
| Detection Method: | Colorimetric |
| Detection Instrument: | Microplate reader (450 nm) |
| Detection Range: | 0.01–50 U/L |
| Sensitivity: | 0.01 U/L |
| Assay Time: | 20 min |
| Research Area: | Glycolysis and lipid metabolism; disulfidptosis |
| Detection Method: | Colorimetric (WST-8 formazan) |
| Wavelength: | 450 nm |
| Detection Range: | 0.01–50 U/L |
| Sensitivity: | 0.01 U/L |
| Intra-assay CV: | 1.9–2.4% (three human serum samples, 20 replicates each) |
| Inter-assay CV: | 5.6–6.1% (three human serum samples, 20 assays in duplicate by three operators) |
| Sample means used for precision: | 2.60, 11.50 and 38.50 U/L |
| Sample Volume (page listing): | 100 µL |
- Colorimetric readout: NADPH reduces WST-8 to orange formazan, read at 450 nm
- Detection range 0.01–50 U/L with 0.01 U/L sensitivity
- Validated sample types: serum, plasma and animal tissue homogenates
- Intra-assay CV 1.9–2.4% and inter-assay CV 5.6–6.1% across three human serum samples
- Kit includes extracting solution, substrate, chromogenic agent, buffer solution and standard powder, 96-well microplate and plate sealers
- Approximately 20 min assay time on a microplate reader
Under the presence of G-6-PD, glucose-6-phosphate is oxidized to 6-PG, and NADP+ is reduced to NADPH. Under the action of the electron coupling reagent 1-MPMS, NADPH reduces WST-8 to form orange formazan, which has a maximum absorption peak at about 450 nm. The formazan generated is proportional to G-6-PD activity in the sample.
- Measuring G-6-PD activity in serum or plasma from human, rat, rabbit and cynomolgus monkey samples
- Quantifying G-6-PD activity in mouse liver, rat kidney, rat spleen and mouse brain tissue homogenates
- Enzyme activity profiling in glycolysis and lipid metabolism studies
- Comparing G-6-PD activity across experimental groups in animal tissue studies
- Research into disulfidptosis-related enzyme activity
| Component | Quantity | Storage |
| Reagent 1: Extracting Solution | 50 mL × 2 vials | -20°C, 12 months |
| Reagent 2: Substrate | 1.5 mL × 2 vials | -20°C, 12 months |
| Reagent 3: Chromogenic Agent | 1.5 mL × 2 vials | -20°C, 12 months, protect from light |
| Reagent 4: Buffer Solution | 4 mL × 1 vial | -20°C, 12 months |
| Reagent 5: Standard Powder | 1 vial | -20°C, 12 months, protect from light |
| Microplate | 96 wells | No special storage requirement |
| Plate Sealer | 2 pieces | Not applicable |
| Sample Layout Sheet | 1 piece | Not applicable |
- Microplate reader capable of 450 nm absorbance
- Pipettor and pipette tips
- 37°C water bath or incubator
- Centrifuge, including capacity for 10,000 × g at 4°C
- Ice
- Double-distilled water
- Phosphate-buffered saline (PBS, 0.01 mol/L, pH 7.4) for tissue washing
- Dounce homogenizer, kept at 4°C for tissue samples
- Protein quantification method for tissue supernatant (for per-gram-protein normalization)
Serum/plasma
- Detect directly.
- If not detected on the same day, store serum or plasma at -80 °C for up to one month.
- Recommended dilution factors (for reference only): human serum 1; rat serum 1; rabbit serum 1; cynomolgus monkey 1.
- The diluent is extracting solution (Reagent 1). For other sample types, perform a pretest to confirm the dilution factor.
Tissue
- Harvest the amount of tissue needed for each assay (initial recommendation 20 mg).
- Wash tissue in cold PBS (0.01 M, pH 7.4).
- Homogenize 20 mg tissue in 180 µL Reagent 1 with a Dounce homogenizer at 4 °C.
- Centrifuge at 10000 × g for 10 minutes at 4 °C to remove insoluble material. Collect the supernatant and keep it on ice for detection.
- Determine the protein concentration of the supernatant using a separate protein assay, because results are expressed per gram of tissue protein.
- Recommended dilution factors for 10% homogenates (for reference only): mouse liver 3-5; rat kidney 1; rat spleen 5-10; mouse brain 1.
- The diluent is extracting solution (Reagent 1). For other tissue types, perform a pretest to confirm the dilution factor.
Reagent preparation
- Keep Reagent 2 (substrate) on ice during use. Equilibrate other reagents to room temperature before use.
- Prepare the 5 mmol/L standard by dissolving one vial of Reagent 5 in 720 µL of double-distilled water. Prepare on the day of use.
- Prepare the 500 µmol/L standard: mix 1 µL of 5 mmol/L standard with 9 µL of buffer solution (Reagent 4) per well. Prepare on the day of use.
- Prepare the sample working solution: per well, mix 25 µL of substrate (Reagent 2) with 25 µL of chromogenic agent (Reagent 3), 50 µL total. Prepare on the day of use.
- Prepare the control working solution: per well, mix 25 µL of chromogenic agent (Reagent 3) with 25 µL of buffer solution (Reagent 4), 50 µL total. Prepare on the day of use.
- Prepare the standard curve by diluting the 500 µmol/L standard with Reagent 1 as follows: 0 µmol/L (0 µL standard, 200 µL Reagent 1); 50 µmol/L (20 µL, 180 µL); 100 µmol/L (40 µL, 160 µL); 150 µmol/L (60 µL, 140 µL); 250 µmol/L (100 µL, 100 µL); 350 µmol/L (140 µL, 60 µL); 400 µmol/L (160 µL, 40 µL); 500 µmol/L (200 µL, 0 µL). Discard working standards after use.
Serum/plasma: G-6-PD activity (U/L)=(ΔA-b)÷a÷T×f; Tissue: G-6-PD activity (U/gprot)=(ΔA-b)÷a÷T×f÷C pr
- ΔA = OD(sample) − OD(control).
- a and b are the slope and intercept of the standard curve y=ax+b, where y is the absolute OD value (blank-subtracted mean OD of each standard) and x is standard concentration in µmol/L.
- T is the reaction time, 10 min.
- f is the dilution factor of the sample before testing. Multiply by f to correct for any sample dilution.
- C pr is the protein concentration of the sample in g prot/L (tissue calculation only).
- Unit definition: 1 unit is the amount of enzyme in 1 L of serum or plasma (or in 1 g of tissue protein) that produces 1 µmol NADPH per minute at 37 °C.
- Units: serum and plasma results are in U/L; tissue results are in U/gprot.
What sample types can be used with this kit?
The kit is intended for serum, plasma, and animal tissue samples. Serum and plasma are tested directly, and tissue must first be homogenized in Reagent 1 and clarified by centrifugation.
How should samples and reagents be stored?
Kit components are stored at -20 °C for 12 months, and the chromogenic agent and standard powder must be protected from light. Serum and plasma not tested on the same day can be stored at -80 °C for up to one month.
What should I do if my sample is outside the detection range?
The detection range is 0.01-50 U/L. If the concentration is not within the detection range, perform an extra dilution or concentration of the sample.
What key points should I follow to avoid errors?
Pour out part of the extracting solution when it is taken to avoid contamination. Prevent bubbles when transferring supernatant into the microplate.
What instrument settings and assay time are used?
Read optical density at 450 nm on a microplate reader after mixing for 5 s and incubating at 37 °C for 10 min. The product page lists the assay time as 20 min.
| Storage: | Kit components at -20°C, with the chromogenic agent and standard powder protected from light |
| Shelf Life: | 12 months |
Notes
- For research use only; do not use in clinical diagnosis.
- Keep substrate on ice during use and equilibrate the other reagents to room temperature before use.
- Prepare standard solutions and sample working solution fresh on the day of use and discard leftover working standard dilutions.
- Do not mix reagents from different kits.
- When removing extracting solution, pour out part of it to avoid contamination, and avoid bubbles when transferring supernatant into the microplate.