Description
Cell Mitochondrial Complex I (NADH-CoQ Reductase) Activity Assay Kit
The Cell Mitochondrial Complex I (NADH-CoQ Reductase) Activity Assay Kit (MAES0464) measures Complex I activity in cell samples by monitoring NADH oxidation as the change in absorbance at 340 nm. Complex I catalyzes the transfer of electrons from NADH to ubiquinone in the electron transport chain. The kit is used to quantify Complex I activity in cell lysates for studies of mitochondrial function and cellular energy metabolism. For Research Use Only.
Mitochondrial complex I, also called NADH-CoQ reductase, catalyzes the reaction of NADH with ubiquinone substrate to generate NAD+ and reduced ubiquinone. This reaction connects NADH oxidation to ubiquinone reduction inside mitochondria, so the activity of the enzyme reflects the mitochondrial electron transfer function that this kit is designed to quantify in cell samples.
Researchers measure complex I activity to study mitochondrial function and energy metabolism, which are the research areas listed for this kit. The intended use is to measure complex I (NADH-CoQ reductase) activity in cell samples. The product is supplied for research use only and is not intended for clinical diagnosis.
The assay follows the change in optical density at 340 nm. Cell homogenate is mixed with a reaction working solution prepared from substrate A and substrate B working solutions, with substrate B dissolved in anhydrous ethanol. The initial absorbance (A1) is read at 340 nm, the mixture is incubated at 37°C for 5 min, and the absorbance is read again (A2). The decrease in OD, ΔA = A1 − A2, is used in the activity calculation. The readout is colorimetric and is performed on a UV microplate reader.
The kit is intended for cell samples, and the manual lists recommended dilution factors of 1 for Jurkat, CHO, HL-60, HeLa, 293T and Molt-4 cells. Adherent or suspension cells are harvested, washed with PBS (0.01 M, pH 7.4), and homogenized in extraction solution containing protease inhibitor. The manual recommends using fresh sample for detection and keeping the supernatant on ice until use. Other sample types require a pretest to confirm the dilution factor.
Performance data in the manual include an analytical sensitivity of 2.85 U/L and a detection range of 2.85–119.04 U/L. Intra-assay CV was 3.0–4.0%, and inter-assay CV was 5.0–8.0%, using three Molt-4 cell samples with mean values of 30, 60 and 90 U/L. Average recovery was 99%, with individual recoveries of 95%, 97% and 105% at expected concentrations of 20, 50 and 100 U/L, respectively.
| Product name: | Cell Mitochondrial Complex I (NADH-CoQ Reductase) Activity Assay Kit |
| SKU: | MAES0464 |
| Sizes: | 48 Tests, 96 Tests |
| Sample type: | Cell samples (adherent or suspension) |
| Detection method: | Colorimetric; change in absorbance at 340 nm |
| Detection wavelength: | 340 nm |
| Assay time: | 5-minute reaction at 37 °C after reaction working solution is added |
| Intended use: | Research Use Only |
| Sensitivity: | 2.85 U/L |
| Detection range: | 2.85-119.04 U/L |
| Intra-assay CV: | 3.0-4.0% |
| Inter-assay CV: | 5.0-8.0% |
| Intra-assay precision (mean, %CV): | 30.00 U/L, 3.0%; 60.00 U/L, 3.8%; 90.00 U/L, 4.0% |
| Inter-assay precision (mean, %CV): | 30.00 U/L, 5.0%; 60.00 U/L, 6.5%; 90.00 U/L, 8.0% |
| Recovery (average): | 99% |
| Recovery by level (expected, observed U/L, recovery %): | 20, 19.0, 95%; 50, 48.5, 97%; 100, 105.0, 105% |
| Sample volume per well: | 20 µL cell homogenate |
| Reaction volume per well: | 200 µL reaction working solution added to 20 µL sample |
- Colorimetric readout: change in absorbance at 340 nm over a 5-minute reaction at 37 °C
- Sample type: adherent or suspension cell samples
- Detection range 2.85-119.04 U/L; sensitivity 2.85 U/L
- Intra-assay CV 3.0-4.0%; inter-assay CV 5.0-8.0%
- Kit sizes: 48 Tests and 96 Tests; stored at -20 °C for 12 months
Mitochondrial Complex I catalyzes the oxidation of NADH to NAD+. Complex I activity is determined by measuring the change in optical density (ΔOD) at 340 nm over a 5-minute incubation at 37 °C, which reflects consumption of the NADH substrate. Activity is calculated from ΔA, the reaction volume, sample volume, reaction time, the NADH molar extinction coefficient (6600 L/(mol•cm)), the optical path (0.7 cm), the sample protein concentration, and the sample dilution factor. One unit is defined as the amount of Complex I that hydrolyzes substrate to produce 1 µmol of product per minute at 37 °C per gram of mitochondrial protein.
- Measuring mitochondrial Complex I activity in cell samples as part of electron transport chain studies
- Investigating cellular energy metabolism by quantifying Complex I enzyme activity
- Comparing Complex I activity across cell types (sample types listed in the manual include Jurkat, CHO, HL-60, HeLa, 293T, and Molt-4 cells)
| Component | Quantity | Storage |
| Extraction Solution | 48 Tests: 25 mL ×1 vial; 96 Tests: 50 mL ×1 vial | -20 °C, 12 months |
| Protease Inhibitor | 48 Tests: 0.8 mL ×1 vial; 96 Tests: 0.8 mL ×2 vials | -20 °C, 12 months, protect from light |
| Buffer Solution | 48 Tests: 15 mL ×1 vial; 96 Tests: 15 mL ×2 vials | -20 °C, 12 months, protect from light |
| Substrate A (powder) | 48 Tests: 1 vial; 96 Tests: 2 vials | -20 °C, 12 months, protect from light |
| Substrate B (powder) | 48 Tests: 1 vial; 96 Tests: 2 vials | -20 °C, 12 months, protect from light |
| 96-well UV microplate | 48 Tests: 1 plate; 96 Tests: 1 plate | No special requirement |
| Plate sealer | 48 Tests: 2 pieces; 96 Tests: 2 pieces | No special requirement |
| Sample layout sheet | 48 Tests: 1 sheet; 96 Tests: 1 sheet | No special requirement |
- Centrifuge capable of 10000×g and 4 °C
- 37 °C incubator
- Microplate reader with 340 nm detection
- Anhydrous ethanol (AR grade)
- PBS (0.01 M, pH 7.4) for cell washing
- Ultrasonic cell disruptor with 4 °C cooling
- Double-distilled water
- Ice
- Protein concentration assay for measuring lysate protein (not included in the kit)
Cells (adherent or suspension)
- Harvest the number of cells needed for each assay. The initial recommendation is 1 × 10^6 cells.
- Wash cells with PBS (0.01 M, pH 7.4).
- Homogenize 1 × 10^6 cells in 200 µL extraction solution and 4 µL protease inhibitor using an ultrasonic cell disruptor (4°C, 200 W, 5 s per pulse, 10 s interval, repeated 15 times).
- Centrifuge at 10000 × g for 3 min at 4°C to remove insoluble material. Collect the supernatant and keep it on ice for detection.
- Determine the protein concentration of the supernatant using a separate protein concentration assay. The result is required as C_pr (g protein/L) in the calculation.
- Dilution: the recommended dilution factor is 1 for Jurkat, CHO, HL-60, HeLa, 293T and Molt-4 cells. The diluent is extraction solution. For other sample types, perform a pretest to confirm the dilution factor.
Reagent preparation
- Equilibrate all reagents to room temperature before use.
- Substrate A solution: dissolve one vial of Substrate A in 150 µL double-distilled water and mix well. Store at -20 °C for 7 days, protected from light.
- Substrate A working solution: prepare fresh before the assay. Example: mix 5 µL Substrate A solution with 500 µL Buffer Solution (505 µL total). Store at 2-8 °C for up to 12 h, protected from light.
- Substrate B working solution: dissolve one vial of Substrate B in 4 mL anhydrous ethanol and shake until the solution turns a yellowish clear liquid. Store at 2-8 °C for 48 h, protected from light, or aliquot and store at -20 °C for 7 days, protected from light.
- Reaction working solution: prepare fresh on the day of use. Example: mix 5 µL Substrate B working solution with 295 µL Substrate A working solution (300 µL total). Keep on ice and use within 1 h.
Mitochondrial complex I activity (U/gprot) = ΔA_Sample ÷ (6600 × 0.7) ÷ T × V1 ÷ V2 × f ÷ C_pr × 10^6
- ΔA_Sample: change in OD of the sample, A1 − A2 (A1 read before incubation, A2 read after 5 min at 37°C at 340 nm).
- 6600: molar extinction coefficient of NADH, L/(mol·cm).
- 0.7: optical path, cm.
- T: reaction time, 5 min.
- V1: volume of the reaction system, 0.22 mL.
- V2: volume of the sample, 0.02 mL (20 µL).
- f: dilution factor of the sample before testing. Multiply by f; use f = 1 for undiluted samples.
- C_pr: protein concentration of the sample, g protein/L.
- 10^6: conversion factor, 1 mol = 10^6 µmol.
- Definition: 1 unit is the amount of mitochondrial complex I in 1 g of mitochondrial protein that hydrolyzes the substrate to produce 1 µmol of product per minute at 37°C.
- Worked example in the manual: ΔA = 0.672 − 0.66 = 0.012 for 20 µL HeLa cell homogenate with C_pr = 1.64 g protein/L, giving 3.48 U/gprot.
What sample volume is used in the assay?
Add 20 µL of sample to each sample well. The manual specifies cell homogenate as the sample type, prepared from 1 × 10^6 cells in 200 µL extraction solution.
How should the kit components be stored?
Store the kit components at -20°C for 12 months, protected from light. Substrate A solution can be stored at -20°C for 7 days protected from light, and substrate A working solution at 2–8°C for 12 h. Reaction working solution should be prepared on the spot and kept on ice for use within 1 h.
What should I do if my results are out of range?
The detection range is 2.85–119.04 U/L. If the concentration is outside this range, an extra dilution or concentration should be applied to the sample. If the OD value decreases by more than 0.3 within 3 min during measurement, the sample should be diluted.
What are the key points for obtaining reliable results?
Make sure substrate B working solution is completely dissolved, extending oscillation or ultrasound time if needed. Use fresh samples, and measure no more than 8 sample wells at the same time. Equilibrate all reagents to room temperature before use.
What instrument settings are required?
Read absorbance at 340 nm with a microplate reader, using a UV microplate. Incubate the reaction at 37°C. A centrifuge is required for sample preparation.
How long does the assay take?
The assay time is 30 min. The reaction is incubated at 37°C for 5 min between the two OD readings at 340 nm.
| Storage: | -20°C; protect light-sensitive components from light |
| Shelf Life: | 12 months |