The Rat Myelin Basic Protein (MBP) ELISA Kit is specially designed to provide precise and accurate measurement of myelin basic protein levels in rat serum, plasma, and cell culture supernatants. This kit is known for its high sensitivity and specificity, delivering reliable and consistent results for various research purposes.Myelin basic protein is a key component of the myelin sheath in the nervous system, playing a crucial role in maintaining the structure and function of nerve fibers.
Dysregulation of MBP levels has been linked to various neurological disorders such as multiple sclerosis and other demyelinating diseases.By using the Rat Myelin Basic Protein (MBP) ELISA Kit, researchers can gain valuable insights into the pathophysiology of these conditions and potentially develop new therapeutic strategies. Its ease of use and robust performance make it an invaluable tool for studying the role of myelin basic protein in neurological health and disease.
Product Name:
Rat Myelin Basic Protein (MBP) ELISA Kit
Product Code:
RTDL00655
Size:
96 Assays
Target:
MBP
Synonyms:
Myelin A1 protein, Myelin membrane encephalitogenic protein
Detection Method:
Sandwich
Range:
0.156-10ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-99
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.