Description
Succinate Assay Kit (BA0149)
The Succinate Assay Kit (SKU: BA0149) provides a simple, one-step method for measuring succinate (succinic acid) in food, beverage, agricultural and other biological samples. Succinate is found in all plant and animal tissues, is an intermediate in the citric acid cycle and plays an important role in intracellular energy generation. In this assay succinate is converted to pyruvate, which reacts with specific reagents and a dye to form a coloured product. The colour intensity at 570 nm, or the fluorescence intensity at 530/585 nm, is directly proportional to the succinate concentration in the sample. The room-temperature format requires no 37 C heater, uses 20 uL of sample and is readily automated for high-throughput use.
The Succinate Assay Kit (BA0149) measures succinate on a standard microplate reader, either colorimetrically (10–400 µM) or fluorometrically (2–40 µM), from a 20 µL sample. Succinate is converted to pyruvate, which reacts with a dye to form a coloured product read at 570 nm or a fluorescent product read at 530/585 nm, in a single 30-minute room-temperature step. Specified sample types: food, beverage, agricultural and other biological samples.
Use the colorimetric readout (10–400 µM) when succinate is relatively abundant, such as in many food and beverage samples. Use the fluorometric readout (2–40 µM, a five-fold lower floor) for low-succinate samples. Both modes use 20 µL of sample. Dilute samples above the top of the range.
An enzymatic kit gives a single-analyte succinate concentration in 96 wells within about 30 minutes, without a mass spectrometer. LC-MS measures succinate alongside other TCA-cycle intermediates and can use isotope tracing, but it needs specialist instrumentation and method development. Use the kit for targeted, high-throughput comparisons, and LC-MS when you need a metabolite panel.
Succinate is a TCA-cycle intermediate oxidised to fumarate by succinate dehydrogenase (mitochondrial complex II). It accumulates when tissue is ischaemic or SDH is impaired, which makes it a common readout in ischaemia–reperfusion, organ-preservation, hypoxia and cancer-metabolism studies, and in fermentation and food-quality testing.
| BA0149 colorimetric | BA0149 fluorometric | LC-MS | |
| Range | 10–400 µM | 2–40 µM | Method-dependent |
| Sample volume | 20 µL | 20 µL | Method-dependent |
| Time | 30 min RT | 30 min RT | Hours incl. prep |
| Instrument | Plate reader (absorbance) | Plate reader (fluorescence) | LC-MS/MS |
| Multiplexed metabolites | No | No | Yes |
| Product Name: | Succinate Assay Kit (BA0149) |
| SKU: | BA0149 |
| Detection Method: | Colorimetric / Fluorometric |
| Detection Range: | Colorimetric 10 - 400 uM; fluorometric 2 - 40 uM succinate |
| Sample Type: | Food, beverage, agricultural products and other biological samples |
| Species Reactivity: | All |
| Assay Time: | 30 minutes |
| Kit Size: | 100 Assays |
| Equipment Required: | Microplate reader |
| Storage: | Store all components at -20 C upon receipt. |
| Shelf Life: | 6 months after receipt |
| Shipping: | Gel Pack |
A one-step enzymatic assay for the quantitative colorimetric or fluorometric determination of succinate. Succinate is converted to pyruvate, which reacts with a dye to form a coloured product at 570 nm or a fluorescent product at 530/585 nm. The room-temperature procedure adds a single working reagent and reads after 30 min.
- Fast and sensitive, using 20 uL of sample
- Linear detection range: colorimetric 10 - 400 uM, fluorometric 2 - 40 uM succinate
- Convenient room-temperature assay requiring no 37 C heater
- High-throughput 96-well plate format
- Direct measurement of succinate in food, beverage, agricultural products and other biological samples
Note: The below protocol is a sample protocol. Protocols are specific to each batch/lot. For the correct instructions please follow the protocol included in your kit.
| Step | Procedure |
| 1 | Reconstitute PEP by adding 120 uL water and mixing thoroughly. Store reconstituted PEP at -20 C and use within one month. |
| 2 | Prepare samples: assay clear and slightly coloured samples directly; homogenise solid samples in water and filter or centrifuge (5 min at 14,000 rpm). |
| 3 | Colorimetric assay uses an internal standard. Add 20 uL of each sample to two wells and 20 uL water to a further well. Prepare 1 mM succinate standard (20 uL of 20 mM in 380 uL water); add 5 uL to the sample-plus-standard wells and 5 uL water to the sample-alone and water wells. |
| 4 | Prepare Working Reagent per well by mixing 85 uL Assay Buffer, 1 uL Enzyme Mix, 1 uL Cosubstrate, 1 uL PEP and 1 uL Dye Reagent. Add 80 uL to each well, tap to mix and incubate for 30 min at room temperature. |
| 5 | Read optical density at 570 nm (550-585 nm). |
| 6 | For the fluorometric assay, prepare 40, 24, 12 and 0 uM standards, transfer 20 uL standards and samples into a black plate, add 80 uL Working Reagent and read fluorescence at 530/585 nm after 30 min. |
Colorimetric: [Succinate] = ((R_SAMPLE - R_WATER) / (R_STANDARD - R_SAMPLE)) x 250 x n (uM), where R values are the optical densities of the sample, water and sample-plus-standard, and n is the dilution factor. Fluorometric: determine the slope from the standards and calculate [Succinate] = ((R_SAMPLE - R_WATER) / Slope) x n (uM). One mM succinate equals 11.7 mg/dL or 117 ppm.
| Component | Quantity | Storage |
| Assay Buffer | 10 mL | -20 C |
| Enzyme Mix | 120 uL | -20 C |
| Cosubstrate | 120 uL | -20 C |
| PEP | Dried | -20 C |
| Dye Reagent | 120 uL | -20 C |
| Standard (20 mM Succinate) | 500 uL | -20 C |
What sample types is the Succinate Assay Kit validated for?
The specification lists food, beverage, agricultural and other biological samples. Serum and plasma are not listed in the specification. Contact technical support before running those matrices.
Should I choose the colorimetric or fluorometric protocol?
Use the colorimetric protocol when succinate is expected between 10 and 400 µM. Use the fluorometric protocol for lower concentrations, between 2 and 40 µM. Both protocols need 20 µL of sample per well and a standard microplate reader. Run a pilot on representative samples if the expected concentration is uncertain.
How long does the succinate assay take?
After sample preparation, the reaction runs for 30 minutes at room temperature. Then you read the plate on an absorbance or fluorescence microplate reader. A kit contains enough reagents for 100 assays. Store all components at −20 °C on receipt.
Is succinate the same analyte as succinic acid?
Yes. Succinic acid is a dicarboxylic acid (pKa values ≈ 4.2 and 5.6), so at neutral pH it exists almost entirely as the succinate dianion. Assay Genie's Succinic Acid Colorimetric Assay Kit (MAES0489) measures the same analyte. BA0149 adds a fluorometric option for lower concentrations.
Does this kit measure succinate dehydrogenase activity?
No. BA0149 measures succinate concentration, which is the metabolite itself. Succinate dehydrogenase (complex II) oxidises succinate to fumarate. To measure that enzyme's activity, use the Succinate Dehydrogenase (SDH) Activity Assay Kit (MAES0229). To quantify SDH subunit protein, use the SDHA or SDHB ELISA kits.
| Sun, K., et al. | First Experience with Hypothermic Oxygenated Perfusion in Human Uteri: Feasibility and Metabolic Characterization | Journal of Clinical Medicine 2026 | View Citation |