MAES0005 typical standard curve - Uric Acid (UA) Fluorometric Assay Kit
Key Features

Uric Acid (UA) Fluorometric Assay Kit (MAES0005)

SKU MAES0005
Product Type Assay
Detection Method Fluorometric
Instrument Microplate Reader
Sample Type Serum, Plasma, Urine
Sensitivity 0.03 umol/L
Range 0.03-15 umol/L
Assay Time 40 min
Shipping Gel Pack
Research Area Oxidative Stress
Research Area ATP and Related
Research Area Diabetes & Obesity
Research Area Clinical Chemistry
€615

Request a formal quotation

A PDF quote with its own quote number, ready for your purchasing office to raise a PO against. Back within 4 working hours, from a PhD scientist — not an autoresponder.

  • PO-ready PDFQuote number, VAT, Incoterms
  • Volume pricing appliedDiscounts from 2 kits up
  • Valid 30 daysPrices held while you get sign-off
Anything we should know? optional

No account needed. We use your details only to prepare and send this quote. If it's urgent, say so in the box and we'll come back the same working day.

Global Shipping: 80+ Countries
White Glove Service: Available upon request
Batch Consistency: Contact Sales
Distributors: 60+ Countries

Description

system_update_altDatasheetsystem_update_altMSDS

Uric Acid (UA) Fluorometric Assay Kit

The Uric Acid (UA) Fluorometric Assay Kit (MAES0005) quantifies uric acid in serum, plasma, urine, and animal tissue samples. Uricase converts uric acid to H2O2, which is then detected with a peroxidase-coupled fluorogenic probe. Researchers use the kit to measure uric acid across purine metabolism and renal and liver function studies on a standard fluorescence microplate reader.

Uric acid (UA) is the analyte this kit quantifies in serum, plasma, urine, and animal tissue samples. For liquid samples, results are reported in µmol/L. For tissue homogenates, results are reported in µmol/g protein, which normalizes the value to tissue protein content. The kit provides a 20 µmol/L uric acid standard, which is diluted with buffer solution to build the calibration curve. In the detection principle, uric acid is the substrate for uricase, the enzyme that starts the signal chain.

The product is listed under the Liver and Renal Function research area, which places UA measurement within studies of hepatic and renal physiology. The sample types named in the dilution guidance include human serum, human urine, and human hydrothorax fluid. They also include rat serum, rat urine, rat liver, rat kidney, and rat lung homogenates, and rabbit and porcine serum. This range lets researchers compare UA levels across these matrices in animal and human-derived samples.

The method is a fluorometric enzymatic reaction. Uricase converts uric acid into allantoin, carbon dioxide, and hydrogen peroxide. In the presence of peroxidase, the hydrogen peroxide oxidizes a non-fluorescent probe into a fluorescent substance. The working solution combines buffer solution, probe solution, and two enzyme reagents, and it is prepared on the day of use and protected from light. After samples or standards are mixed with working solution, the plate is mixed for 5 s and incubated at 37°C for 30 min. Fluorescence is then read at an excitation of 535 nm and an emission of 587 nm in a fluorescence microplate reader.

Serum, plasma, and other liquid samples are tested directly. If serum or plasma is not tested on the same day, it can be stored at -80°C for one month. Tissue samples are harvested in the amount needed for each assay, with an initial recommendation of 20 mg, and washed in cold PBS (0.01 M, pH 7.4). The tissue is homogenized in buffer solution at 4°C, then centrifuged to remove insoluble material. The supernatant is kept on ice for detection. Urine is diluted in buffer solution before testing. Reagents must be stored at -20°C and the probe solution must be shaded from light.

Reported sensitivity is 0.03 µmol/L, and the detection range is 0.03-15 µmol/L. The product page lists intra-assay CV as 1.5% and inter-assay CV as 7.2%. The manual reports intra-assay CVs of 1.2% to 2.0% and inter-assay CVs of 6.8% to 7.9% across three human serum samples. Recovery averaged 101%, and individual recovery rates ranged from 99% to 104%. The product page lists an assay time of 40 min.