Aldose Reductase Inhibitor Screening Kit Colorimetric Assay (ARMA00126)
Aldose reductase (AR) catalyzes the NADPH-dependent reduction of an aldose (a sugar molecule containing a terminal aldehyde) to the corresponding alcohol. It is an important member of the polyol pathway, an alternative route for glucose metabolism that operates at high glucose levels. Accumulation of sorbitol due to AR activity, which is not sufficiently converted to fructose by sorbitol dehydrogenase, is believed to be the main cause of diabetic complications such as cataracts, retinopathy, neuropathy and nephropathy, as sorbitol cannot diffuse across cell membranes and causes osmotic stress in cells.
Several aldose reductase inhibitors have shown promise in managing diabetic complications in clinical settings. Assay Genie's Aldose Reductase Inhibitor Screening Kit is a simple colorimetric means of evaluating potential inhibitors of aldose reductase, including determination of IC50 values for test compounds.
| Product Name: | Aldose Reductase Inhibitor Screening Kit Colorimetric Assay |
| SKU: | ARMA00126 |
| Size: | 100 Wells |
| Product Type: | Inhibitor Screening Kit |
| Detection Method: | Colorimetric (OD 340 nm), measured kinetically |
| Application: | Screening of potential aldose reductase inhibitors, including IC50 determination |
| Inhibitor Control: | Epalrestat |
| Storage: | Store unopened kit at -20°C. Bring all components to room temperature before using. |
- Screens potential inhibitors of aldose reductase, the first enzyme of the polyol pathway.
- Colorimetric detection (OD 340 nm), measured kinetically.
- Suitable for IC50 determination of test compounds using serial dilutions.
- Includes Epalrestat as a known aldose reductase inhibitor control.
- Simple 96-well plate format with enzyme, background and solvent controls.
- 100 wells.
Aldose reductase reduces D-xylose to xylitol while converting NADPH to NADP+. The oxidation of NADPH is followed at 340 nm as a decrease in absorbance. Inhibitors suppress the decrease in absorbance.
| Component | Size | Cap Colour |
| Assay Buffer | 35 ml | NM |
| DTT | 200 µl | Green |
| D-Xylose | 1 ml | Red |
| Aldose Reductase | 2 x Lyophilized | Purple |
| NADPH | Lyophilized | Blue |
| Epalrestat | 20 µl | Yellow |
*Note: The below protocol is a sample protocol. Protocols are specific to each batch/lot. For the correct instructions please follow the protocol included in your kit.
| Step | Protocol |
| 1. | Plate Reader Setup: Turn on the plate reader and set it to 37°C. |
| 2. | Assay Buffer: Prepare 200 µl of Assay Buffer for each well to be run and add 2 µl of diluted DTT per 200 µl. Prepare only the amount needed, as DTT is unstable and oxidizes over time. Include 1 or 2 Test Compound wells, a Background Control, Enzyme Control and Solvent Control for each compound tested. |
| 3. | Test Compounds (TC): Dissolve test compounds at 100X the final desired concentration in an appropriate solvent. Transfer 10 µl to a tube containing 40 µl Assay Buffer, then add 10 µl of each diluted compound to the test wells. |
| 4. | Controls: Inhibitor Control (IC): add 10 µl of diluted Epalrestat (2 µl Epalrestat in 198 µl Assay Buffer). Enzyme Control (EC): add 10 µl Assay Buffer. Background Control (BC): add 100 µl Assay Buffer. Solvent Control (SC, optional): add 10 µl of diluted solvent (10 µl solvent in 40 µl Assay Buffer). |
| 5. | NADPH Solution: Add 60 µl of NADPH solution (3.3 µl NADPH stock + 57 µl Assay Buffer per well) to all wells. |
| 6. | Aldose Reductase Solution: Add 90 µl of Aldose Reductase solution (2 µl Aldose Reductase + 88 µl Assay Buffer containing DTT per well) to all Test Compound, Inhibitor Control, Enzyme Control and Solvent Control wells. Shake for several seconds and incubate at 37°C for 15 minutes to allow inhibitor binding. |
| 7. | Initiate Reaction: Add 40 µl of D-Xylose solution (4 µl D-Xylose + 36 µl Assay Buffer per well, pre-warmed at 37°C for 5-10 minutes) to all wells. All wells should now contain 200 µl. |
| 8. | Measurement: Immediately place the plate in the plate reader and measure absorbance at 340 nm in kinetic mode for 30-60 minutes. |
| 9. | Calculation: Calculate the slope of the linear portion of each curve and subtract the Background Control slope. % Inhibition = [1 – (|SlopeTC – SlopeBC| / |SlopeEC – SlopeBC|)] × 100. Use the Solvent Control slope in place of the Enzyme Control slope if they differ noticeably. For IC50 values, perform a serial dilution of each test compound. |