Description
Calcium (Ca) Colorimetric Assay Kit
The Calcium (Ca) Colorimetric Assay Kit (MAES0091) quantifies calcium in serum, plasma, urine, and tissue homogenates. Calcium in the sample binds Methyl Thymol Blue (MTB) under alkaline conditions to form a blue complex, which is read at 610 nm on a microplate reader. The kit provides a 20-minute workflow with a detection range of 0.07-1.2 mmol/L, suited to researchers measuring calcium levels across biological sample types.
The Calcium (Ca) Colorimetric Assay Kit quantifies calcium ion content in biological samples. The product is listed under the Inorganic Ions series and the Plant Physiology research area, with Ca as its synonym. Calcium is the analyte measured directly in the sample, so the readout reflects total calcium content rather than a specific binding partner or enzyme activity. For tissue samples, results are normalized to protein concentration and expressed per gram of protein. The kit is for research use only and is not intended for clinical diagnosis.
Publications that cite this kit span bone regeneration, wearable monitoring of sweat and physiological signals, and neuroexcitotoxicity in hippocampal neurons. The cited work includes a multifunctional bionic periosteum with ion sustained-release for bone regeneration, a modular wearable body net for monitoring sweat and physiological signals, and a Capsicum annuum seed extract tested in HT22 hippocampal neurons. The manual's example analysis also covers dog serum, rat heart tissue homogenate, and a plant tissue homogenate, which shows use across animal and plant sample types.
The assay is colorimetric. Calcium ions in the sample bind Methyl Thymol Blue in alkaline solution to form a blue complex. The product page gives a specific absorption peak at 715 nm, while calcium content is calculated from the optical density at 610 nm. Each well receives 10 µL of standard or sample and 250 µL of working solution, is mixed for 30 s on the microplate reader, and stands for 5 min at room temperature before reading at 610 nm. Working solution 1 is used for serum, plasma, and other liquid samples. Working solution 2, which contains clarificant, is used for tissue homogenates. Total assay time is 20 min.
Serum, plasma, and urine are tested directly. Serum or plasma that is not tested on the same day can be stored at -80°C for one month. Tissue is washed in cold PBS (0.01 M, pH 7.4), homogenized at 4°C in deionized water at 20 mg per 180 µL, and centrifuged at 10000×g for 10 min to remove insoluble material. The supernatant is kept on ice, and its protein concentration must be determined separately. Severe hemolysis, jaundice, and lipidemia affect results, and calcium contamination must be avoided throughout the assay.
Analytical sensitivity is 0.07 mmol/L. It was determined by adding two standard deviations to the mean OD of the zero standard assayed 20 times and calculating the corresponding concentration. The detection range is 0.07-1.2 mmol/L. The product page reports an intra-assay CV of 4.7% and an inter-assay CV of 8.5%. The manual's replicate data for three human serum samples show intra-assay CVs of 4.6-4.8% and inter-assay CVs of 8.4-8.6%. Recovery at low, middle, and high concentrations ranged from 97% to 101%, with an average of 99%.
| Product name: | Calcium (Ca) Colorimetric Assay Kit |
| SKU: | MAES0091 |
| Size: | 96 Assays |
| Sample types: | Serum, plasma, urine, tissue homogenate |
| Detection method: | Colorimetric (MTB dye) |
| Wavelength: | 610 nm (optimum); microplate reader 600-620 nm |
| Assay time: | 20 min |
| Detection range: | 0.07-1.2 mmol/L |
| Sensitivity: | 0.07 mmol/L |
| Intra-assay CV: | 4.7% |
| Inter-assay CV: | 8.5% |
| Recovery: | 99% |
| Detection method: | Colorimetric |
| Target: | Calcium (Ca2+) |
| Detection wavelength: | 610 nm |
| Sample volume: | 10 µL |
| Assay time: | 20 min |
| Sensitivity: | 0.07 mmol/L |
| Detection range: | 0.07-1.2 mmol/L |
| Intra-assay CV: | 4.7% |
| Inter-assay CV: | 8.5% |
| Recovery: | 99% |
| Research area: | Inorganic ions |
- Colorimetric MTB method read at 610 nm on a microplate reader
- Detection range 0.07-1.2 mmol/L; analytical sensitivity 0.07 mmol/L
- Assay time of 20 min with 10 µL sample input
- Validated intra-assay CV 4.7% and inter-assay CV 8.5%
- Mean recovery of 99%
- Supplied as a 96 Assay kit with 2.5 mmol/L calcium standard
Calcium ions in the sample bind Methyl Thymol Blue (MTB) in alkaline solution to form a blue complex. The complex is read by measuring the optical density (OD) at 610 nm, and calcium content is calculated from a linear standard curve (y = ax + b) using blank-corrected absorbance.
- Quantification of calcium in serum, plasma, and urine samples
- Measurement of calcium content in tissue homogenates, normalized to protein concentration
- Studies of calcium signaling in mammalian development and reproduction
- Research on calcium dynamics in cancer and bone-related models
| Component | Quantity | Storage |
| MTB Reagent | 10 mL × 1 vial | 2-8°C, protect from light |
| Alkali Reagent | 20 mL × 1 vial | 2-8°C |
| Clarificant | 1 mL × 1 vial | 2-8°C |
| Calcium Standard (2.5 mmol/L) | 10 mL × 1 vial | 2-8°C |
| Microplate | 96 wells | No special requirement |
- Microplate reader capable of reading 600-620 nm (610 nm optimum)
- Micropipettor
- Centrifuge (capable of 10000×g for tissue samples)
- Incubator
- Vortex mixer
- Deionized water
- 0.01 M PBS, pH 7.4 (for washing tissue)
- Dounce homogenizer (for tissue samples)
Serum, plasma, and urine
- Detect serum, plasma, and urine samples directly.
- If serum or plasma is not detected on the same day, store it at -80°C for up to one month.
- Dilute samples as needed using the dilution factors in the table below. The manual lists deionized water as the diluent.
- For sample types not listed, perform a pretest to confirm the dilution factor.
Tissue
- Harvest the amount of tissue needed for each assay. The initial recommendation is 20 mg.
- Wash the tissue in cold PBS (0.01 M, pH 7.4).
- Homogenize 20 mg of tissue in 180 µL of deionized water with a dounce homogenizer at 4°C. Use deionized water as the homogenization medium to avoid calcium contamination.
- Centrifuge at 10000×g for 10 min to remove insoluble material. Collect the supernatant and keep it on ice for detection.
- Determine the protein concentration of the supernatant separately, using the protein assay kit recommended in the manual, for use in the calculation.
- Dilute the 20% tissue homogenate with a dilution factor of 1 (no additional dilution).
Reagent preparation
- Equilibrate all reagents to room temperature before use.
- Preheat the clarificant at 37°C until clear before use, as it solidifies at 2-8°C.
- Serum, plasma, and urine: prepare working solution 1 fresh on the day of use (90 µL MTB reagent + 180 µL alkali reagent per well, scaled to the number of wells).
- Tissue homogenates: prepare working solution 2 fresh (90 µL MTB reagent + 180 µL alkali reagent + 9 µL clarificant per well, scaled to the number of wells).
- Standard curve: dilute the 2.5 mmol/L standard with deionized water to 0, 0.2, 0.3, 0.4, 0.6, 0.8, 1.0, and 1.2 mmol/L (for example, 0 to 240 µL of standard brought to 500 µL total with deionized water). Prepare fresh and discard after use.
Sample preparation
- Serum, plasma, and urine: assay directly. If not assayed the same day, store serum or plasma at -80°C for up to one month.
- Tissue: wash tissue in cold 0.01 M PBS, pH 7.4. Homogenize 20 mg tissue in 180 µL deionized water using a Dounce homogenizer at 4°C.
- Centrifuge the homogenate at 10000×g for 10 min. Collect the supernatant and keep it on ice.
- Determine the protein concentration of the supernatant separately for tissue normalization.
- Dilute samples as needed. Recommended dilution factors (reference only): dog serum 2-3; human serum 3-6; mouse serum 3-6; human urine 4-8; 20% animal tissue homogenate 1. Pretest other sample types to confirm dilution.
Assay procedure
- Add 10 µL of each standard or sample to the corresponding microplate wells.
- Add 250 µL of working solution 1 (serum, plasma, urine) or working solution 2 (tissue homogenate) to each well.
- Mix thoroughly for 30 s on the microplate reader, then stand for 5 min at room temperature.
- Measure the OD at 610 nm.
Calculation
- Average the duplicate readings for each standard. Subtract the mean blank (standard ①, 0 mmol/L) OD from all readings to obtain absolute OD values.
- Plot absolute OD (y) against standard concentration (x) and fit a linear standard curve (y = ax + b).
- Liquid samples: calcium (mmol/L) = (ΔA610 − b) ÷ a × f, where f is the dilution factor and ΔA = absolute OD of sample (OD sample − OD blank).
- Tissue samples: calcium (mmol/g protein) = (ΔA610 − b) ÷ a × f ÷ Cpr, where Cpr is protein concentration in g protein/L.
Serum (plasma) and other liquid samples: Calcium content (mmol/L) = (ΔA610 - b) ÷ a × f Tissue and cell samples: Calcium content (mmol/gprot) = (ΔA610 - b) ÷ a × f ÷ C_pr Standard curve: y = ax + b, where y is the absolute OD and x is the standard concentration in mmol/L.
- a: slope of the standard curve.
- b: intercept of the standard curve.
- ΔA610: absolute OD, calculated as OD(sample) minus OD(blank).
- f: dilution factor of the sample before testing. For undiluted samples, f = 1. For diluted samples, multiply the result by the dilution factor.
- C_pr: concentration of protein in the sample, in gprot/L. Tissue results are expressed in mmol/gprot.
- Subtract the mean OD of the blank (Standard #1, 0 mmol/L) from all standard readings to obtain absolute OD values before plotting the standard curve.
- Example: with y = 0.1298x + 0.0018, a sample OD of 0.308, and a blank OD of 0.221, a dog serum sample diluted 2-fold gives (0.308 - 0.221 - 0.0018) ÷ 0.1298 × 2 = 1.31 mmol/L.
What sample types can be tested, and which dilutions should I use?
The kit is validated for serum, plasma, urine, and tissue samples. The manual's recommended dilution factors are 2-3 for dog serum, 3-6 for human and mouse serum, 4-8 for human urine, and 1 for 20% animal tissue homogenate. For other sample types, perform a pretest to confirm the dilution factor.
How should samples and reagents be stored?
Store the kit reagents at 2-8°C for 12 months, and keep the MTB reagent protected from light. Serum or plasma that is not tested on the same day can be stored at -80°C for one month.
What should I do if my result is outside the detection range?
The detection range is 0.07-1.2 mmol/L. If the concentration is not within this range, dilute or concentrate the sample as needed and retest it.
What can interfere with the assay?
Severe hemolysis, jaundice, and lipidemia affect results. Avoid calcium contamination throughout the procedure, and use deionized water as the homogenization medium when preparing tissue homogenates.
What instrument settings are required?
Read the optical density at 610 nm on a microplate reader with a 600-620 nm filter range. The optimum wavelength is 610 nm.
How long does the assay take?
The assay time is 20 min. After adding working solution, mix for 30 s, stand for 5 min at room temperature, and then read the plate at 610 nm.
| Storage: | 2-8°C |
| Light sensitivity: | MTB reagent: store with shading from light |
| Shelf life: | 12 months |
| Shipping: | Gel pack |
| Sample storage: | Serum or plasma: -80°C for up to one month |
Notes
- For Research Use Only. Not for use in clinical diagnosis or therapeutic procedures.
- Avoid calcium contamination throughout the assay, including when preparing tissue homogenates. Use deionized water as the homogenization medium.
- Severe hemolysis, jaundice, and lipidemia can affect results.
- Do not mix reagents from different kit lots.
- Prepare working solutions and standard dilutions fresh on the day of use; discard unused working standards.
- If a sample falls outside the detection range, dilute or concentrate it and re-assay.
- Wear a lab coat and gloves during handling.
Calcium (Ca) Colorimetric Assay Kit (MAES0091) has been used in 1 publication. Newest first.
- 2025Biochanin a modulates steroidogenesis and cellular metabolism in human granulosa cells through TAS2Rs activation: a spotlight on ovarian functionView publication ↗Luongo, F.P., et al. · Reproductive Biology and Endocrinology