Description
Formate Colorimetric Assay Kit
The Formate Colorimetric Assay Kit (MAES0095) quantifies formate in serum, plasma, and animal tissue samples. Formate is detected through formic acid dehydrogenase (FDH)-catalyzed NADH generation, which is converted by PMS and WST-8 into a yellow product read at 450 nm on a microplate reader. Researchers use the kit to measure formate content in biological samples for studies of formic acid metabolism and exposure. For Research Use Only.
The Formate Colorimetric Assay Kit (MAES0095) quantifies formate content in biological samples. Its detection chemistry starts with formic acid dehydrogenase (FDH), which catalyzes the reaction of formic acid with NAD+ to produce NADH. The kit then converts that NADH into a measurable colored signal, so the optical readout reflects formate present in the sample. The sources describe the chemistry but do not describe the broader metabolic role of formate in physiology.
The kit's intended use is to measure formate content in serum, plasma and animal tissue samples. The sources list human, dog, mouse, horse and porcine serum, rat plasma, and rat liver, kidney, spleen and brain homogenates in their dilution guidance, and they also show human urine in an example analysis. The sources do not describe specific disease models or physiological states. The kit is labeled for research use only and must not be used for clinical diagnosis.
Standards and samples (50 µL each) are combined with a measuring working solution containing extracting solution, substrate A and substrate B working solutions, and chromogenic agent. After a 5 s mix on the microplate reader, the plate is incubated at 37°C for 30 min. NADH, acting through PMS, transfers electrons to WST-8 to form a yellow product with a characteristic absorption peak at 450 nm. Because sample NAD(P)H causes background interference, a control well is included in the measurement to eliminate it.
Serum and plasma are centrifuged through a 10 KD ultrafiltration tube at 12000×g for 10 min at 4°C, and the filtrate is assayed. Tissue is washed in cold PBS (0.01 M, pH 7.4), homogenized at 4°C, clarified at 10000×g, and ultrafiltered in the same way. Reagents are stored at -20°C for 12 months, with the substrate and chromogenic components protected from light. Substrate working solutions and measuring working solution are prepared on the day of use and kept on ice, and working standard dilutions are discarded after use.
Analytical sensitivity is 8.20 µmol/L, and the detection range is 8.20-800 µmol/L. Intra-assay CV was 3.1% to 3.6% and inter-assay CV was 3.8% to 4.1% across three human serum samples. Recovery was 99% to 101% at three expected concentrations of 125, 362 and 554 µmol/L. The total assay time is 1 h 10 min.
| Product Name: | Formate Colorimetric Assay Kit |
| SKU: | MAES0095 |
| Product Size: | 96 Assays |
| Sample Type: | Serum, plasma, animal tissue |
| Detection Method: | Colorimetric |
| Instrument: | Microplate reader (450 nm) |
| Assay Time: | 1 h 10 min |
| Intended Use: | Measurement of formate content in serum, plasma, and animal tissue samples (Research Use Only) |
| Sensitivity: | 8.20 µmol/L |
| Detection Range: | 8.20-800 µmol/L |
| Sample Volume: | 50 µL |
| Intra-assay CV: | 3.3% |
| Inter-assay CV: | 3.9% |
| Recovery Rate: | 100% |
| Detection Wavelength: | 450 nm |
| Assay Time: | 1 h 10 min |
- Colorimetric endpoint readout at 450 nm on a microplate reader
- Detection range 8.20-800 µmol/L with analytical sensitivity of 8.20 µmol/L
- Compatible with serum, plasma, and animal tissue homogenates
- Sample volume 50 µL per well; assay time 1 h 10 min
- Intra-assay CV 3.3% and inter-assay CV 3.9%
- Supplied as 96 Assays per kit; stored at -20°C for 12 months
Formic acid dehydrogenase (FDH) catalyzes the reaction of formic acid with NAD+ to produce NADH. NADH, under the action of PMS, transfers electrons to WST-8 to produce a yellow product with a characteristic absorption peak at 450 nm. NAD(P)H present in the sample contributes background signal, so a control well is included in the measurement to correct for this interference.
- Quantification of formate in serum and plasma samples
- Measurement of formate in animal tissue homogenates
- Studies of formic acid metabolism and clearance
- Research on formate accumulation associated with methanol exposure or formaldehyde exposure
- Investigation of formic acid use and effects in livestock feed research
| Component | Quantity | Storage |
| Extracting Solution (Reagent 1) | 60 mL x 2 vials | -20°C, 12 months, protect from light |
| Substrate A (Reagent 2), powder | 2 vials | -20°C, 12 months, protect from light |
| Substrate B (Reagent 3), powder | 1 vial | -20°C, 12 months, protect from light |
| Chromogenic Agent (Reagent 4) | 1.5 mL x 2 vials | -20°C, 12 months, protect from light |
| 10 mmol/L Standard (Reagent 5) | 1.0 mL x 1 vial | -20°C, 12 months |
| Microplate | 96 wells | No requirement |
- Microplate reader capable of 450 nm absorbance
- Pipettor
- Water bath set to 37°C
- Centrifuge (including 4°C refrigerated centrifugation at 10000-12000 x g)
- Double distilled water
- 10 kDa ultrafiltration tubes
- Dounce homogenizer
- Ice
- Cold PBS (0.01 M, pH 7.4) for tissue washing
Serum/plasma
- Centrifuge the sample through a 10 KD ultrafiltration tube at 12000×g for 10 min at 4°C.
- Collect the filtrate and use it for detection.
- Dilution factor is 1 for human serum, dog serum, mouse serum, horse serum, porcine serum and rat plasma.
- Use extracting solution as the diluent. For other sample types, run a pretest to confirm the dilution factor.
Animal tissue
- Harvest the amount of tissue needed for each assay (initial recommendation 20 mg).
- Wash the tissue in cold PBS (0.01 M, pH 7.4).
- Homogenize 20 mg of tissue in 180 µL of extracting solution with a dounce homogenizer at 4°C.
- Centrifuge at 10000×g for 10 min at 4°C to remove insoluble material.
- Collect the supernatant and centrifuge it through a 10 KD ultrafiltration tube at 12000×g for 10 min.
- Take the filtrate for detection.
- Dilution factor is 1 for 10% rat liver, kidney, spleen and brain tissue homogenates.
- Use extracting solution as the diluent. For other tissue types, run a pretest to confirm the dilution factor.
Reagent preparation
- Equilibrate all reagents to room temperature before use.
- Dissolve one vial of Substrate A with 170 µL of Extracting Solution; mix well and keep on ice during use. Prepare on the day of use.
- Dissolve one vial of Substrate B with 300 µL of Extracting Solution; mix well and keep on ice during use. Prepare on the day of use.
- Prepare measuring working solution fresh before the assay: 40 µL Extracting Solution, 5 µL Substrate A working solution, 5 µL Substrate B working solution, and 50 µL Chromogenic Agent (100 µL total). Keep on ice and protected from light.
- Prepare standards fresh by diluting the 10 mmol/L standard in double distilled water: 100 µmol/L (10 µL standard + 990 µL water), 200 µmol/L (20 µL + 980 µL), 300 µmol/L (30 µL + 970 µL), 400 µmol/L (40 µL + 960 µL), 500 µmol/L (50 µL + 950 µL), 600 µmol/L (60 µL + 940 µL), 800 µmol/L (80 µL + 920 µL). Use 0 µmol/L (1000 µL water) as the blank. Discard working standards after use.
Sample preparation
- Serum and plasma: centrifuge through a 10 kDa ultrafiltration tube at 12000 x g for 10 min at 4°C. Use the filtrate for the assay.
- Tissue: weigh tissue (initial recommendation 20 mg) and wash in cold PBS (0.01 M, pH 7.4).
- Homogenize tissue in extracting solution (manual example: 20 mg in 180 µL) with a Dounce homogenizer at 4°C.
- Centrifuge at 10000 x g for 10 min at 4°C. Collect the supernatant and filter through a 10 kDa ultrafiltration tube at 12000 x g for 10 min. Use the filtrate for the assay.
- Dilute samples as needed; the manual lists a dilution factor of 1 for the reference sample types (serum, plasma, and 10% tissue homogenates). Pretest other sample types to confirm dilution.
Assay procedure
- Add 50 µL of sample to sample wells and 50 µL of each standard to standard wells.
- Add 50 µL of measuring working solution to sample and standard wells.
- Mix for 5 s on the microplate reader and incubate at 37°C for 30 min.
- Read absorbance at 450 nm. Average duplicate readings and subtract the blank (0 µmol/L standard) mean OD to obtain absoluted OD values.
- Plot absoluted OD against concentration and fit a linear standard curve (y = ax + b). Calculate formate content as described in the Calculation section.
Serum/plasma: Formate content (µmol/L) = (ΔA - b) ÷ a × f Tissue: Formate content (µmol/g wet weight) = (ΔA - b) ÷ a × f × V ÷ W
- ΔA = OD(sample) - OD(blank), where the blank OD is the OD of the standard at concentration 0 µmol/L.
- a and b are the slope and intercept of the standard curve y = ax + b, plotted with absorbance (absoluted OD) on the y-axis and concentration on the x-axis.
- f is the dilution factor of the sample before testing. For the listed samples, f = 1.
- V is the volume of extraction solution used in tissue homogenization, stated as 0.9 mL = 0.0009 L.
- W is the weight of the sample, stated as 0.1 g.
- Results are reported in µmol/L for serum and plasma, and in µmol/g wet weight for tissue.
Which sample types can I test with this kit?
The kit measures formate in serum, plasma and animal tissue samples. The dilution table lists human, dog, mouse, horse and porcine serum, rat plasma, and 10% rat liver, kidney, spleen and brain homogenates.
How should I store the kit and prepared reagents?
Kit components should be stored at -20°C for 12 months, with the substrate and chromogenic components protected from light. Substrate A and B working solutions and the measuring working solution should be prepared on the spot and kept on ice, protected from light.
What should I do if my sample is outside the detection range?
The detection range is 8.20-800 µmol/L. If the concentration is not within this range, take an extra dilution or concentration of the sample.
What causes background interference, and how is it handled?
NAD(P)H present in the sample itself causes background interference. Set a control well during measurement to eliminate this interference. Avoid bubbles when adding samples, and break any bubbles before reading.
What instrument settings are required?
Read the optical density at 450 nm on a microplate reader. Mix the plate for 5 s, then incubate at 37°C for 30 min before reading. Keep the highest standard OD within 1.5-2.0 when the sample color allows.
How long does the assay take?
The assay time is 1 h 10 min. Color continues to deepen with longer incubation, so keep the incubation consistent between runs.
| Storage: | -20°C, protect from light |
| Shelf Life: | 12 months |
| Shipping: | Gel pack |
Notes
- For Research Use Only. Not for use in clinical diagnosis or therapeutic procedures.
- Avoid bubbles when adding samples; break bubbles before reading.
- Prepare substrate and measuring working solutions fresh on the day of use and keep on ice and protected from light.
- Prepare standards fresh and discard working dilutions after use.
- Color continues to develop with incubation time; adjust incubation so the highest standard OD falls within 1.5-2.0 when sample color allows.
- Do not mix reagents from different kits. Centrifuge small reagent volumes before use.
- Include the control well to correct for NAD(P)H background in samples.
- If a sample falls outside the detection range, dilute or concentrate the sample and re-assay.