Description
Cell Glutathione Peroxidase (GPX) Activity Assay Kit
The Cell Glutathione Peroxidase (GPX) Activity Assay Kit (MAES0457) measures GPX activity in cell samples by colorimetric detection at 340 nm. GPX reduces peroxides using glutathione and protects cells from lipid peroxidation caused by free radicals. Researchers use the kit to quantify GPX activity in oxidative stress and ferroptosis studies.
Glutathione peroxidase (GPX) is an enzyme that protects cells from free radical damage caused by hydrogen peroxide decomposition. Lipid components of cells are prone to reacting with free radicals, which leads to lipid peroxidation. GPX uses glutathione to reduce peroxides to alcohols, thereby preventing cell damage. Its activity therefore reflects how effectively a cell neutralizes peroxide-driven injury, making it a central component of cellular redox defense.
This kit is used in research on oxidative stress and ferroptosis, the two research areas listed for the product. Both fields depend on understanding how cells handle peroxides and lipid peroxidation, so a direct measure of GPX activity provides a functional readout alongside other markers. Measuring activity in cell lysates allows investigators to compare enzyme function across treatments, cell lines, or experimental conditions under matched handling. The kit is labeled for research use only and must not be used for clinical diagnosis.
GPX catalyzes its substrate, and the product consumes the reducing agent after enzyme reagents are added. The reducing agent has a maximum absorbance at 340 nm, so GPX activity is calculated from the change in absorbance at 340 nm. Cell lysate (20 µL) is combined with reaction working solution (180 µL) and oxidant working solution (40 µL) in a UV microplate. The first OD reading at 340 nm (A1) is taken immediately after oxidant addition, the plate is incubated at room temperature for 10 min, and a second reading (A2) is taken. The total assay time is 30 min, and the method is colorimetric.
Suitable samples are cells, including adherent or suspension cultures. The kit's listed reference cell types are HL-60, 293T, CHO, 4T1, and Jurkat cells, with a recommended dilution factor of 1 for each. For each assay, cells are washed with PBS (0.01 M, pH 7.4), then homogenized in cell lysis buffer with stabilizer. Lysates are kept on ice and centrifuged to remove insoluble material, and the supernatant is used for detection. Lysate must be kept on ice during use. Reagents are stored at -20°C, protected from light, and the kit remains valid for 12 months.
Performance data reported for this kit include an analytical sensitivity of 3.22 U/L and a detection range of 3.22 to 446.95 U/L. Intra-assay precision is 1.2 to 3.1% CV, and inter-assay precision is 0.4 to 3.4% CV. In a recovery test across high, middle, and low concentration samples, recovery ranged from 96.33% to 100%. Appendix data were generated with HL-60 cells, the reference cell type for the precision and recovery studies.
| Product Name: | Cell Glutathione Peroxidase (GPX) Activity Assay Kit |
| SKU: | MAES0457 |
| Sample Type: | Cell |
| Detection Method: | Colorimetric |
| Detection Instrument: | Microplate reader (340 nm) |
| Detection Range: | 3.22-446.95 U/L |
| Sensitivity: | 3.22 U/L |
| Sample Volume: | 20 µL (cell homogenate) |
| Assay Time: | 30 min |
| Research Area: | Oxidative Stress, Ferroptosis |
| Wavelength: | 340 nm |
| Detection Range: | 3.22-446.95 U/L |
| Sensitivity: | 3.22 U/L |
| Intra-assay CV: | 1.2-3.1% |
| Inter-assay CV: | 0.4-3.4% |
| Recovery: | 96.33-100% |
| Sample Volume: | 20 µL |
| Assay Time: | 30 min |
| Shelf Life: | 12 months |
- Colorimetric measurement of GPX activity at 340 nm on a standard microplate reader
- Detection range 3.22-446.95 U/L with analytical sensitivity of 3.22 U/L
- Intra-assay CV 1.2-3.1% and inter-assay CV 0.4-3.4%
- Requires 20 µL of cell sample per well; assay time 30 min
- Results calculated as U/g protein, normalized to cell protein concentration
- Research Use Only
GPX catalyzes a reaction in which the product consumes a reducing agent, and the enzyme reagents are added to drive the reaction. The reducing agent has a maximum absorbance at 340 nm. GPX activity is calculated from the change in absorbance at 340 nm over the reaction time, after subtracting the blank well change.
- Measuring GPX activity in cell lysates in oxidative stress studies
- Assessing antioxidant enzyme activity in ferroptosis research
- Comparing GPX activity across cell lines, such as HL-60, 293T, CHO, 4T1, and Jurkat cells, using the recommended dilution factor of 1 for these samples
- Evaluating how experimental treatments affect cellular glutathione-dependent peroxide detoxification
| Component | Storage |
| Reagent 1 Buffer Solution | -20°C, 12 months |
| Reagent 2 Substrate Powder | -20°C, 12 months, protect from light |
| Reagent 3 Enzyme Reagent | -20°C, 12 months, protect from light |
| Reagent 4 Oxidant | -20°C, 12 months, protect from light |
| Reagent 5 Reductant Powder | -20°C, 12 months, protect from light |
| Reagent 6 Accelerant | -20°C, 12 months, protect from light |
| Reagent 7 Stabilizer | -20°C, 12 months, protect from light |
| Reagent 8 Cell Lysis Buffer | -20°C, 12 months |
| UV Microplate, 96 wells | No special requirement |
| Plate Sealer | — |
| Sample Layout Sheet | — |
- Microplate reader capable of reading at 340 nm
- Incubator
- Ice
- Centrifuge capable of 10000×g at 4°C
- PBS (0.01 M, pH 7.4) for cell washing
- Double distilled water (for reconstituting substrate and reductant)
- Normal saline (0.9% NaCl) or PBS (0.01 M, pH 7.4) for sample dilution
- Protein concentration assay for the cell lysate supernatant
- Pipettes and sterile tips
- Cell counting method to harvest 1×106 cells per sample
Cells (adherent or suspension)
- Harvest the number of cells needed for each assay (initial recommendation: 1×10^6 cells).
- Wash cells with PBS (0.01 M, pH 7.4).
- Homogenize 1×10^6 cells in 99 µL cell lysis buffer and 1 µL stabilizer.
- Crack on ice for 10 minutes, mixing well every 5 minutes.
- Centrifuge at 10000×g for 10 min at 4°C to remove insoluble material.
- Collect the supernatant and keep it on ice for detection.
- Determine the protein concentration of the supernatant separately, as it is required for the calculation.
- Dilution: the recommended dilution factor is 1 for HL-60 (1.9×10^6 cells), 293T (1.4×10^6 cells), CHO (2×10^6 cells), 4T1 (1.9×10^6 cells), and Jurkat (2.03×10^6 cells) cells. The diluent is cell lysis buffer. For other sample types, perform a pretest to confirm the dilution factor.
Reagent preparation
- Equilibrate all reagents to room temperature before use.
- Substrate working solution: dissolve one vial of substrate powder in 250 µL double distilled water and mix well. Store at -20°C for 2 days, protected from light.
- Reductant working solution: dissolve one vial of reductant powder in 500 µL double distilled water and mix well. Store at -20°C for 2 days, protected from light.
- Oxidant working solution: mix 95 µL Buffer Solution with 5 µL Oxidant per 100 µL prepared. Prepare on the day of use, keep on ice, and use within 2 hours.
- Reaction working solution: prepare on the day of use and keep on ice. Example for one run: 825 µL Buffer Solution, 10 µL substrate working solution, 25 µL Enzyme Reagent, 25 µL reductant working solution, and 400 µL Accelerant (1285 µL total). Use within 2 hours.
Sample preparation (cell samples)
- Harvest the number of cells needed for each assay (initial recommendation 1×106 cells).
- Wash cells with PBS (0.01 M, pH 7.4).
- Homogenize 1×106 cells in 99 µL Cell Lysis Buffer and 1 µL Stabilizer. Crack on ice for 10 minutes, mixing well every 5 minutes.
- Centrifuge at 10000×g for 10 min at 4°C to remove insoluble material. Collect the supernatant and keep it on ice.
- Determine the protein concentration of the supernatant with a separate protein assay.
- Dilute samples as needed. Recommended dilution factor is 1 for HL-60, 293T, CHO, 4T1, and Jurkat cells; for other sample types, run a pretest to confirm the dilution factor.
Assay
- Sample well: add 20 µL of sample. Blank well: add 20 µL of Cell Lysis Buffer.
- Add 180 µL of reaction working solution to each well.
- Add 40 µL of oxidant working solution to each well.
- Mix fully on the microplate reader for 5 s. Immediately measure absorbance at 340 nm and record as A1.
- Incubate at room temperature for 10 min, then measure absorbance at 340 nm and record as A2.
- Calculate ΔAsample = A1 − A2 for the sample well and ΔAblank for the blank well.
Calculation
- GPX activity (U/g protein) = (ΔAsample − ΔAblank) ÷ (ε × d) × (Vtotal ÷ Vsample) ÷ T × f ÷ Cpr
- ε = 6.22×10-3 L/µmol/cm (molar extinction coefficient of product at 340 nm); d = 0.6 cm (optical path); Vtotal = 0.24 mL; Vsample = 0.02 mL; T = 10 min; f = sample dilution factor; Cpr = protein concentration of sample (g protein/L).
- One unit is defined as the amount of GPX in 1 g cell protein per minute that converts 1 µmol of substrate to product at 25°C.
GPX activity (U/gprot) = (ΔA_sample − ΔA_blank) ÷ (ε × d) × (V_total ÷ V_sample) ÷ T × f ÷ C_pr
- ΔA_sample = A1 − A2 for the sample well, where A1 is the OD at 340 nm read immediately after adding oxidant working solution and A2 is the OD at 340 nm after 10 min of incubation at room temperature.
- ΔA_blank = A1 − A2 for the blank well (cell lysis buffer in place of sample).
- ε = molar extinction coefficient of the product at 340 nm, 6.22×10^-3 L/µmol/cm.
- d = optical path, 0.6 cm.
- V_total = total reaction volume, 0.24 mL.
- V_sample = sample volume, 0.02 mL.
- T = reaction time, 10 min.
- C_pr = protein concentration of the sample, g protein/L.
- f = dilution factor of the sample before testing (1 for the listed cell types at the recommended dilution).
- Definition: 1 unit is the amount of GPX in 1 g of cell protein that catalyzes the substrate to produce 1 µmol of product per minute at 25°C. Result units are U/gprot.
What sample volume and cell input should I use?
Add 20 µL of cell lysate per sample well, and add 20 µL of cell lysis buffer to the blank well. The initial recommendation is to harvest 1×10^6 cells per assay, homogenized in 99 µL cell lysis buffer and 1 µL stabilizer.
How should the reagents be stored, and how long do working solutions last?
The kit components are stored at -20°C for 12 months, with the substrate, enzyme reagent, oxidant, reductant, and stabilizer protected from light. Substrate and reductant working solutions are stored at -20°C for 2 days, protected from light, while the oxidant and reaction working solutions must be prepared on the spot, kept on ice, and used within 2 hours.
What should I do if my sample reads out of range?
If the A2 of the sample well is less than 0.2, the sample must be diluted and tested again. The detection range is 3.22 to 446.95 U/L, and if the concentration is not within this range, an extra dilution or concentration should be performed.
What are the key points for an accurate result?
Read A1 immediately after adding oxidant working solution, because the reaction rate is fast. If A1 is not determined in time, the sample may show a low or no value. Equilibrate all reagents to room temperature before use, and centrifuge small reagent volumes before use.
Which instrument settings are required?
Measure optical density at 340 nm using a microplate reader, and use a UV microplate. An incubator is also required for the 10 min room-temperature incubation.
How long does the assay take?
The assay time is 30 min. This includes the 10 min incubation between the two absorbance readings.
| Storage: | -20°C |
| Shelf Life: | 12 months |
| Light Protection: | Store reagents protected from light (shading light) |
| Substrate and Reductant Working Solutions: | Store at -20°C for 2 days, protected from light |
| Oxidant and Reaction Working Solutions: | Prepare fresh, keep on ice, use within 2 hours |
Notes
- For Research Use Only. Not for clinical diagnosis or therapeutic use.
- Wear a lab coat and gloves during use.
- If the A2 of the sample well is less than 0.2, dilute the sample and test again.
- The reaction rate is fast. Record A1 immediately after adding oxidant working solution; delayed reading may give low or no sample values.
- Do not mix reagents from different kits.
- Centrifuge small-volume reagents briefly before use to ensure sufficient volume is recovered.
- Prepare the oxidant and reaction working solutions on the day of use and use within 2 hours.
- Results are sensitive to reagent handling, operation, and environment. Run a pretest for sample types not listed in the protocol.
- If the sample concentration falls outside the detection range, dilute or concentrate the sample as needed.