Description
Cell Hydrogen Sulfide (H2S) Fluorometric Assay Kit
The Cell Hydrogen Sulfide (H2S) Fluorometric Assay Kit (MAES0505) measures intracellular H2S levels in cell samples. A cell-permeant probe enters the cell and emits green fluorescence on reaction with H2S, and signal intensity is read at Ex/Em 490/528 nm. Researchers use the kit to study H2S roles in cellular signal transduction and oxidative stress research. This product is for Research Use Only.
Hydrogen sulfide (H2S) is a gas molecule that functions as a key signal transduction messenger in cells and organisms. It participates in numerous physiological and pathological processes, including vascular relaxation and nerve transmission. Because H2S acts inside cells, its intracellular level is directly relevant to how cells transmit regulatory signals.
Researchers measure intracellular H2S because abnormal concentrations are associated with diseases such as diabetes and liver cirrhosis. Measuring the level can help explore the specific mechanisms by which H2S operates in cellular signal transduction. The same measurements can support early diagnosis, condition monitoring, and evaluation of treatment effects. This kit is positioned within the oxidative stress research area.
The kit uses a fluorescent probe that enters the cell and produces green fluorescence upon reaction with H2S. The intracellular H2S level is determined by measuring fluorescence intensity. The probe is diluted 50-200 times in the supplied buffer, with 50 times as the initial recommended dilution factor, and the working solution is prepared protected from light and used within 1 day. Cells are incubated at 37°C for 30-60 minutes protected from light, then washed twice with PBS (0.01 M, pH 7.4). Readout uses excitation/emission at 490/528 nm on a fluorescence microplate reader, a fluorescence microscope, or a flow cytometer set to FITC.
Suitable samples are cells, including suspension cells and adherent cells cultured in multiwell plates or Petri dishes. Suspension cells are used at a minimum of 1×10^5 cells per 2 mL tube, and adherent cells at a minimum of 5×10^4 cells per well of a 24-well plate. Cells should be washed with PBS before probe incubation. Reagents should be equilibrated to 25°C before use. The probe should be aliquoted and stored at -20°C protected from light, and repeated freeze/thaw cycles should be avoided.
Precision is reported as inter-assay coefficient of variation (CV) of 1.0-9.5% and intra-assay CV of 0.9-3.0%. The recommended sample volume is 0.1 mL, and the total assay time is 1 h 30 min. Sensitivity differs among detection instruments, so the probe working solution dilution factor should be adjusted in a pre-experiment against the instrument signal. A negative control group is recommended for each experiment.
| Product name: | Cell Hydrogen Sulfide (H2S) Fluorometric Assay Kit |
| Assay Genie SKU: | MAES0505 |
| Intended use: | Measurement of H2S level in cell samples |
| Sample type: | Cell |
| Detection method: | Fluorometric |
| Detection instruments: | Fluorescence microplate reader, fluorescence microscope, flow cytometry |
| Research area: | Oxidative stress |
| Assay time: | 1 h 30 min |
| Sample volume: | 0.1 mL |
| Use: | Research Use Only |
| Detection method: | Fluorometric |
| Excitation/Emission: | 490 nm / 528 nm |
| Intra-assay CV: | 0.9-3.0% |
| Inter-assay CV: | 1.0-9.5% |
| Assay time: | 1 h 30 min |
| Sample type: | Cell |
- Fluorometric detection of intracellular H2S in cells
- Green-fluorescent probe; read at Ex/Em 490/528 nm
- Compatible with fluorescence microplate reader, fluorescence microscope, and flow cytometry (FITC channel)
- Intra-assay CV: 0.9-3.0%; Inter-assay CV: 1.0-9.5%
- Assay time of 1 h 30 min; sample volume of 0.1 mL
The probe enters the cell and produces green fluorescence upon reaction with H2S. Intracellular H2S level is determined by measuring fluorescence intensity at Ex/Em 490/528 nm.
- Investigation of intracellular H2S levels and their role in cellular signal transduction
- Oxidative stress research
- Studies of H2S involvement in vascular relaxation and nerve transmission
- Monitoring of intracellular H2S in cell-based experimental models, including treatment-response studies
| Component | Storage |
| Reagent 1: Buffer Solution | -20°C, 12 months, protect from light |
| Reagent 2: Probe | -20°C, 12 months, protect from light |
| Black microplate | No special requirement |
| Plate sealer | No special requirement |
| Sample layout sheet | No special requirement |
- Fluorescence microplate reader, fluorescence microscope, or flow cytometer
- PBS (0.01 M, pH 7.4)
- Black 96-well microplate (for microplate reader readout)
- Centrifuge capable of 500 × g
- 37 °C incubator
- Culture medium and standard cell culture consumables
- Pipettes and microcentrifuge tubes (2 mL EP tubes for suspension cells)
Cells (suspension cells)
- Use at least 1×10^5 cells per 2 mL EP tube.
- Wash the cells twice with PBS (0.01 M, pH 7.4). Centrifuge at 500×g for 3 min at room temperature.
- For the negative control group, add 0.2 mL of buffer solution to resuspend the cells. For the sample group, add 0.2 mL of probe working solution to resuspend the cells. Adjust working solution concentration and incubation time to the experimental situation.
- Incubate at 37°C for 30-60 min protected from light.
- Wash the cells twice with PBS (0.01 M, pH 7.4). Centrifuge at 500×g for 3 min at room temperature.
- For a fluorescence microplate reader, add 0.1 mL of cell suspension to a 96-well black microplate. For flow cytometry, perform direct detection. For fluorescence microscopy, make slides of cells and photograph for observation.
Cells (adherent cells)
- For a 24-well plate, seed at least 5×10^4 cells per well.
- Remove the medium and wash the cells twice with PBS (0.01 M, pH 7.4).
- For the negative control group, add 0.3 mL of buffer solution. For the sample group, add 0.3 mL of probe working solution.
- Incubate at 37°C for 30-60 min protected from light. Adjust working solution concentration and incubation time to the experimental situation.
- Wash the cells twice with PBS (0.01 M, pH 7.4).
- For a fluorescence microplate reader, after cell digestion add PBS (0.01 M, pH 7.4) to resuspend the cells, then add 0.1 mL of cell suspension to a 96-well black microplate. For flow cytometry, after cell digestion add PBS (0.01 M, pH 7.4) to resuspend the cells. For fluorescence microscopy, take photographs for observation.
Reagent preparation
- Equilibrate all reagents to 25°C before use.
- Aliquot the probe and store at -20°C protected from light; avoid repeated freeze/thaw cycles.
- Prepare probe working solution by diluting the probe 50-200× in Buffer Solution. The initial recommended dilution factor is 50×.
- Prepare the probe working solution on the spot, protected from light, and use within 1 day.
- Working solution volumes: 0.1 mL/well (96-well plate), 0.3 mL/well (24-well plate), 0.2 mL/tube (2 mL EP tube, suspension cells).
Suspension cells
- Use at least 1 × 105 cells per 2 mL EP tube. Wash cells twice with PBS (0.01 M, pH 7.4). Centrifuge at 500 × g for 3 min at room temperature.
- Set up a negative control for each experiment. Add 0.2 mL of Buffer Solution to the negative control cells. Add 0.2 mL of probe working solution to sample cells.
- Incubate at 37°C for 30-60 min protected from light. Adjust working solution concentration and incubation time as needed.
- Wash cells twice with PBS (0.01 M, pH 7.4). Centrifuge at 500 × g for 3 min at room temperature.
- Microplate reader: add 0.1 mL of cell suspension to a black 96-well plate and read fluorescence. Flow cytometry: direct detection (FITC channel). Fluorescence microscope: prepare slides and image.
Adherent cells
- For a 24-well plate, seed at least 5 × 104 cells per well. Remove medium and wash cells twice with PBS (0.01 M, pH 7.4).
- Set up a negative control for each experiment. Add 0.3 mL of Buffer Solution to negative control wells. Add 0.3 mL of probe working solution to sample wells.
- Incubate at 37°C for 30-60 min protected from light. Adjust working solution concentration and incubation time as needed.
- Wash cells twice with PBS (0.01 M, pH 7.4).
- Microplate reader: after cell digestion, add PBS (0.01 M, pH 7.4) to resuspend cells, then add 0.1 mL of cell suspension to a black 96-well plate and read fluorescence. Flow cytometry: after digestion, resuspend in PBS (0.01 M, pH 7.4). Fluorescence microscope: image cells.
What sample types can be used with this kit?
The kit measures H2S in cell samples. Both suspension cells and adherent cells are supported, with minimum cell numbers of 1×10^5 per 2 mL tube and 5×10^4 per well of a 24-well plate, respectively.
How should the probe and kit components be stored?
The buffer solution and probe are stored at -20°C for 12 months protected from light. Aliquot the probe, store it at -20°C protected from light, and avoid repeated freeze/thaw cycles. Prepare the probe working solution fresh and use it within 1 day.
What is the recommended probe dilution and how do I adjust it?
Dilute the probe 50-200 times in buffer solution, with 50 times as the initial recommended dilution factor. Sensitivity differs among instruments, so run a pre-experiment to adjust the dilution factor against your instrument signal.
What instrument settings should I use?
Use excitation/emission of 490/528 nm on a fluorescence microplate reader or fluorescence microscope. For flow cytometry, use the FITC channel. Incubate cells at 37°C for 30-60 min protected from light.
What is the total assay time and what precision can be expected?
The assay time is 1 h 30 min. Reported inter-assay CV is 1.0-9.5% and intra-assay CV is 0.9-3.0%. A negative control group is recommended for each experiment.
Is this kit intended for clinical diagnosis?
No. The kit is for research use only and is not intended for clinical diagnosis. Read the instructions completely and adjust instruments before the experiment.
| Storage: | -20°C |
| Storage duration: | 12 months |
| Light sensitivity: | Protect from light (shading light) |
| Valid period: | 12 months |
| Probe handling: | Aliquot; avoid repeated freeze/thaw cycles |
| Working solution stability: | Use within 1 day; prepare protected from light |
| Reagent equilibration: | 25°C before use |
Notes
- For Research Use Only. Not for use in clinical diagnosis or therapeutic procedures.
- Do not mix reagents from different kits.
- Centrifuge small-volume reagents before use to ensure sufficient recovery.
- Store reagents strictly at the stated conditions and protect the probe from light.
- Avoid repeated freeze/thaw cycles of the probe.
- If the analyte concentration falls outside the detection range, dilute or concentrate the sample.
- Run a pre-test if the sample type is not listed in the manual.
- Wear a lab coat and latex gloves during use.
- Adjust probe working solution dilution and incubation time for the specific instrument sensitivity.