The Human c-kit (SCFR) ELISA Kit is specifically developed for the precise measurement of c-kit (stem cell factor receptor) levels in human serum, plasma, and cell culture supernatants. With its exceptional sensitivity and specificity, this kit ensures accurate and reproducible results, making it an indispensable tool for various research applications. The c-kit receptor, also known as SCFR (stem cell factor receptor), plays a critical role in the regulation of hematopoiesis, melanogenesis, and gametogenesis.
Dysregulation of c-kit signaling has been linked to a variety of diseases, including cancer and various hematological disorders. Therefore, the measurement of c-kit levels is essential for studying the role of this receptor in health and disease, as well as for the development of potential targeted therapies. Trust the Human c-kit (SCFR) ELISA Kit for precise and reliable measurements of c-kit levels, allowing for deeper insights into the biological functions of this essential receptor.
Product Name:
Human c-Kit/SCFR ELISA Kit
SKU:
HUFI00188
Reactivity:
Human
Assay Type:
Sandwich ELISA, Double Antibody
Sensitivity:
0.469 ng/mL
Range:
0.781-50 ng/mL
Sample Type:
Serum, Plasma, Cell Culture Supernatant, Cell or Tissue Lysate, Other Liquid Samples
Storage:
2-8°C for 12 months.
Linearity:
Sample
1
2
3
Serum (n = 5)
85-104%
92-97%
89-105%
EDTA Plasma (n = 5)
92-97%
90-101%
87-96%
Heparin Plasma (n = 5)
85-97%
83-94%
84-100%
Recovery:
Sample
1
2
Serum (n = 5)
91-101
92
EDTA Plasma (n = 5)
87-102
91
Heparin Plasma (n = 5)
90-101
96
Note:The below protocol is a sample protocol. Protocols are specific to each batch/lot. For the correct instructions please follow the protocol included in your kit.
Step
Procedure
1
Reagent & Plate Preparation: Equilibrate reagents and TMB substrate to room temperature. Set standard, test sample and control (zero) wells on the pre-coated plate and record their positions.
2
Primary Incubation: Prepare standards, samples, blanks and load into designated wells. Incubate plate at 37°C for 90 minutes to allow antigen binding.
3
Detection Antibody Binding: Add biotin-labeled detection antibody and incubate at 37°C for 60 minutes.
4
HRP-Streptavidin Binding: Add HRP-Streptavidin (SABC) and incubate at 37°C for 30 minutes.
5
Color Development: Add TMB substrate and incubate in the dark for 10–20 minutes.
6
Stop Reaction & Reading: Add stop solution and measure absorbance at 450 nm immediately.
Sample Type
Protocol
Serum
Allow blood to clot, centrifuge at 1000 × g for 20 minutes, collect supernatant supernatant and store appropriately.
Plasma
Collect using anticoagulant tubes, centrifuge at 1000 × g for 15 minutes at 2–8°C and collect plasma.
Tissue Homogenate
Homogenize tissue in PBS with protease inhibitors, centrifuge and collect supernatant.
Cell Culture Supernatant
Centrifuge at 2500 rpm for 5 minutes and collect clarified supernatant.
Cell Lysate
Lyse cells using lysis buffer with protease inhibitors, centrifuge and collect protein supernatant.
Other Sample Types
For more information about how to process other sample types, (e.g., body fluids, breast milk & more), please contact our Tech Support Team at techsupport@assaygenie.com.
Component
Quantity
Storage
48T
96T
ELISA Microplate (Dismountable)
8×6
8×12
Place the test strips into a sealed foil bag with the desiccant. Store for 1 month at 2-8°C; Store for 12 months at -20°C.
Lyophilized Standard
1 vial
2 vial
Place the standards into a sealed foil bag with the desiccant. Store for 1 month at 2-8°C; Store for 12 months at -20°C.