Description
Mitochondrial ROS Fluorometric Assay Kit
The Mitochondrial ROS Fluorometric Assay Kit (MAES0299) measures mitochondrial superoxide and reactive oxygen species (ROS) in living cells using a probe that localizes to mitochondria and fluoresces red after oxidation. It is designed for fluorescence microplate reader, flow cytometry, and fluorescence microscopy workflows. Researchers use it to assess cellular redox status and to evaluate how compounds affect mitochondrial function and oxidative stress.
Mitochondrial reactive oxygen species (ROS) are reactive molecules generated within mitochondria, and mitochondrial superoxide is one ROS species of this type. This kit is intended to measure superoxide in living cell mitochondrial samples, so the readout reflects an oxidant pool inside intact cells. The kit also detects mitochondrial ROS using a separate excitation wavelength, allowing both signals to be assessed within the same workflow.
The research areas listed for this kit are oxidative stress, ferroptosis, and metabolic diseases, which are the contexts in which mitochondrial ROS measurements are applied. Measuring superoxide directly in living cells allows mitochondrial oxidant production to be assessed in these settings. The kit is labeled for research use only and is not intended for clinical diagnosis.
The fluorescent probe specifically targets mitochondria in living cells and is oxidized by mitochondrial superoxides to produce bright red fluorescence. Superoxide is detected with excitation at 396 nm, while mitochondrial ROS is detected with excitation at 510 nm; both are read at an emission of 610 nm. Readout can be performed on a fluorescence microplate reader, a fluorescence microscope, or a flow cytometer. Cells are incubated with the probe working solution at 37°C for 10 to 60 minutes protected from light, and the total assay time is 1 h 30 min.
The kit is designed for cultured suspension cells and adherent cells, which are washed twice with measuring buffer before probe addition. Probe working solution is diluted in measuring buffer at 1:200 to 1:2000, with 1:2000 as a suitable starting point, and is prepared on the day of use and used within 1 day. The probe is stored at -20°C protected from light, and repeated freeze/thaw cycles should be avoided. Reagents are equilibrated to 25°C before use. Inter-assay precision is 6.2-11.7% CV.
| Product name: | Mitochondrial ROS Fluorometric Assay Kit |
| SKU: | MAES0299 |
| Size: | 96T |
| Sample type: | Cells (adherent or suspension) |
| Detection method: | Fluorometric |
| Detection principle: | Probe oxidized by mitochondrial superoxide to produce bright red fluorescence |
| Excitation/emission (superoxide): | 396 nm / 610 nm |
| Excitation/emission (mitochondrial ROS): | 510 nm / 610 nm |
| Incubation: | 37 °C, 10 to 60 minutes, protected from light |
| Assay time: | 1 h 30 min |
| Compatible instruments: | Fluorescence microplate reader, flow cytometry, fluorescence microscope |
| Research areas: | Oxidative stress, cellular stress, enzymes, ferroptosis, metabolic diseases |
| Use: | Research Use Only |
| Sample type: | Cells (adherent or suspension) |
| Detection method: | Fluorometric |
| Superoxide detection: | Ex 396 nm / Em 610 nm |
| Mitochondrial ROS detection: | Ex 510 nm / Em 610 nm |
| Incubation temperature: | 37 °C |
| Incubation time: | 10 to 60 minutes, protected from light |
| Inter-assay CV: | 6.2% to 11.7% |
| Shelf life: | 12 months |
- Mitochondria-targeted fluorescent probe for live-cell measurements
- Superoxide detection at Ex/Em 396 nm / 610 nm; mitochondrial ROS detection at Ex/Em 510 nm / 610 nm
- Compatible with fluorescence microplate reader, flow cytometry, and fluorescence microscope
- Suitable for adherent and suspension cells
- Incubation at 37 °C for 10 to 60 minutes, protected from light
- Inter-assay CV of 6.2% to 11.7%
The fluorescent probe in this kit targets mitochondria in living cells. Mitochondrial superoxide oxidizes the probe, producing bright red fluorescence. Excitation at 396 nm detects mitochondrial superoxide specifically, and excitation at 510 nm detects mitochondrial ROS. Fluorescence intensity is directly proportional to the level of mitochondrial ROS in the sample.
- Measuring mitochondrial superoxide and ROS levels in living cells
- Assessing cellular redox status
- Evaluating the effects of compounds on mitochondrial function
- Studying oxidative stress in cell models
- Imaging mitochondrial ROS by fluorescence microscopy or flow cytometry
| Component | Quantity | Storage |
| Reagent 1: Buffer Solution | 20 mL x 1 vial | -20 °C, 12 months, protect from light |
| Reagent 2: Probe | 0.08 mL x 1 vial | -20 °C, 12 months, protect from light |
| Black microplate | 96 wells x 1 plate | No special requirement |
| Plate sealer | 2 pieces | Not applicable |
- Fluorescence microplate reader, flow cytometer, or fluorescence microscope
- Incubator set to 37 °C
- Double-distilled water
- Centrifuge for small-volume reagents
- Pipettes and 2 mL microcentrifuge (EP) tubes
- Cell culture medium and washing supplies for removing medium
Suspension cells
- Prepare the cell suspension and set up control and sample tubes (2 mL EP tube). Use at least 2×10^5 cells per tube.
- Remove the medium and wash the cells twice with measuring buffer solution.
- Sample tube: remove the supernatant and add 0.2 mL of probe working solution to the resuspended cells.
- Control tube: remove the supernatant and add 0.2 mL of measuring buffer solution to the resuspended cells.
- Incubate at 37°C for 10 to 60 minutes protected from light. Adjust probe concentration and incubation time according to the experimental situation.
- Prepare measuring buffer on the spot by mixing 9 mL of double distilled water with 1 mL of buffer solution (example for 10 mL). Dilute the probe in measuring buffer at 1:200 to 1:2000, starting at 1:2000, and optimize empirically for cell type and cell number.
Adherent cells
- Set up control and sample tubes, with at least 1×10^5 cells per tube.
- Remove the medium and wash the cells twice with measuring buffer solution.
- Sample tube: add probe working solution. Control tube: add measuring buffer solution.
- Incubate at 37°C for 10 to 60 minutes protected from light. Adjust probe concentration and incubation time according to the experimental situation.
- Use the following volumes of probe working solution or measuring buffer solution per well: 1.5 mL (6-well plate), 0.3 mL (24-well plate), 0.1 mL (96-well plate), and 1.5 mL (35 mm petri dish).
- Prepare measuring buffer on the spot by mixing 9 mL of double distilled water with 1 mL of buffer solution (example for 10 mL). Dilute the probe in measuring buffer at 1:200 to 1:2000, starting at 1:2000, and optimize empirically for cell type and cell number.
Reagent preparation
- Equilibrate all reagents to 25 °C before use.
- Prepare measuring buffer solution on the day of use by mixing 9 mL of double-distilled water with 1 mL of Reagent 1 (Buffer Solution) to make 10 mL. Scale to the number of test wells.
- Prepare probe working solution by diluting Reagent 2 (Probe) in measuring buffer solution. The recommended dilution range is 1:200 to 1:2000, with 1:2000 as a starting point. Optimize dilution empirically for each cell type and cell number.
- Prepare probe working solution fresh, protect from light, and use within 1 day.
- Centrifuge small-volume reagents before use.
Suspension cells
- Prepare the cell suspension. Set up control and sample tubes (2 mL EP tubes) with at least 2 x 10^5 cells per tube.
- Remove the medium and wash cells twice with measuring buffer solution.
- Sample tube: remove supernatant and add 0.2 mL probe working solution. Control tube: remove supernatant and add 0.2 mL measuring buffer solution.
- Incubate at 37 °C for 10 to 60 minutes, protected from light. Adjust probe concentration and incubation time to the experiment.
- Read fluorescence on a microplate reader or analyze by flow cytometry.
Adherent cells
- Set up control and sample wells with at least 1 x 10^5 cells per tube or well. Remove medium and wash twice with measuring buffer solution.
- Sample: add probe working solution. Control: add measuring buffer solution.
- Incubate at 37 °C for 10 to 60 minutes, protected from light. Adjust probe concentration and incubation time to the experiment.
- Read by fluorescence microplate reader or flow cytometry, or observe and photograph by fluorescence microscopy.
Instrument settings
- Superoxide: Ex/Em 396 nm / 610 nm. ROS: Ex/Em 510 nm / 610 nm.
- Flow cytometry: use the Violet 610 or PE-Texas Red channel.
- Fluorescence microscopy: use the Texas Red or RFP filter set.
What sample types can be used with this kit?
The kit measures superoxide in living cell mitochondrial samples. Suspension cells require at least 2×10^5 cells per tube, and adherent cells require at least 1×10^5 cells per tube.
How should the kit components be stored?
The Buffer Solution and Probe are stored at -20°C for 12 months, protected from light. Measuring buffer solution is prepared on the spot and can be stored at -20°C for up to a week protected from light, while probe working solution must be prepared on the spot and used within 1 day.
What should I do if my sample is outside the detection range?
If the concentration of a substance is not within the detection range, an extra dilution or concentration should be applied to the sample. It is recommended to run a pre-test if the sample type is not listed in the manual.
What factors can affect results?
The probe optimal dilution may vary with cell type and cell number, so empirical optimization is required for each condition. Results are closely related to reagent condition, operation, and environment, and repeated freeze/thaw cycles of the probe should be avoided.
What instrument settings should I use?
On a fluorescence microplate reader, use Ex/Em 396 nm/610 nm for superoxide and 510 nm/610 nm for ROS. For flow cytometry, use Violet610 or PE-TexasRed, and for fluorescence microscopy, use the TexasRed or RFP channel.
How long does the assay take?
The total assay time is 1 h 30 min. The probe incubation step is performed at 37°C for 10 to 60 minutes protected from light.
| Storage temperature: | -20 °C |
| Storage duration: | 12 months |
| Light protection: | Protect from light (store shaded) |
| Probe handling: | Avoid repeated freeze/thaw cycles |
| Working solutions: | Prepare probe working solution and measuring buffer on the day of use; use probe working solution within 1 day |
| Equilibration: | Equilibrate reagents to 25 °C before use |
Notes
- Research Use Only. Not for clinical diagnosis or therapeutic use.
- Wear a lab coat and latex gloves during use.
- Optimal probe dilution and incubation time vary with cell type and cell number; optimize empirically.
- If the sample concentration falls outside the detection range, dilute or concentrate the sample.
- Run a pre-test if the sample type is not described in the manual.
- Do not mix reagents from different kits.
- Results depend on reagent handling, operation, and environment. Plan sample quantities in advance because sample consumption is not covered by the kit guarantee.