Description
Blood Ammonia Colorimetric Assay Kit
The Blood Ammonia Colorimetric Assay Kit (MAES0116) quantifies ammonia in serum and plasma samples using a Berthelot reaction read on a microplate reader. Protein precipitation in the sample stops enzymatic ammonia formation and removes most interfering colored substances before color development. The kit is used in liver biomarker, amino acid, and protein metabolism research. It is for Research Use Only.
Blood ammonia is the analyte measured by the Blood Ammonia Colorimetric Assay Kit (MAES0116) in serum and plasma samples. The kit quantifies free ammonia as a concentration in mmol/L. Free ammonia can form in a sample through enzymatic activity, and the source notes that glutamine and peptides are easily hydrolyzed to release ammonia. This makes sample handling a direct determinant of the measured value. Ammonia content in red blood cells is reported as 2.8 times higher than in plasma, so hemolysis must be controlled.
Researchers use this kit in studies of liver and renal function and in studies of amino acids and proteins, the two research areas listed on the product page. The kit is intended to measure blood ammonia content in serum and plasma, and it is labeled for Research Use Only. The manual states that it must not be used in clinical diagnosis. Listed sample sources include human, mouse, rat, dog, and horse serum or plasma.
The assay precipitates blood proteins with a protein precipitator, which destroys enzyme activity and prevents further free ammonia formation in vitro. After centrifugation, the non-protein filtrate is treated by the Berthelot reaction, which forms indigo. Most interfering colored substances are removed in this step. Color depth is proportional to blood ammonia content and is compared with a standard curve. Absorbance is read at 635 nm on a microplate reader after 25 min incubation at 37°C.
Serum and plasma are the supported sample types, and each standard or sample tube receives 100 µL of sample. Recommended dilution factors of 1 apply to human serum, human plasma, mouse serum, rat plasma, dog serum, and horse serum. Hemolysis must be avoided during collection and handling. Samples should be tested promptly. They can be stored at 2–8°C for 2–4 hours or at −20°C for 24 hours, and they should be sealed immediately after sampling to prevent ammonia loss.
The stated analytical sensitivity is 0.01 mmol/L, and the detection range is 0.01–2.5 mmol/L. The product page lists an intra-assay CV of 4.1% and an inter-assay CV of 7.2%. The manual reports intra-assay CVs of 3.8–4.3% and inter-assay CVs of 7.0–7.5% across three human serum samples. Recovery of standards at low, middle, and high concentrations was 101%, 104%, and 104%. The total assay time is 45 min.
| Product Name: | Blood Ammonia Colorimetric Assay Kit |
| SKU: | MAES0116 |
| Product Size: | 96 Assays |
| Sample Type: | Serum, plasma |
| Detection Method: | Colorimetric (Berthelot reaction) |
| Instrument: | Microplate reader, 600–660 nm (635 nm optimal) |
| Assay Time: | 45 min |
| Sample Volume: | 0.1 mL |
| Research Area: | Liver biomarkers, amino acids and proteins |
| Intended Use: | Research Use Only |
| Sensitivity: | 0.01 mmol/L |
| Detection Range: | 0.01–2.5 mmol/L |
| Intra-assay CV: | 4.1% |
| Inter-assay CV: | 7.2% |
| Recovery: | 101–104% |
- Colorimetric Berthelot-based detection of blood ammonia in serum and plasma
- Detection range 0.01–2.5 mmol/L with 0.01 mmol/L sensitivity
- Assay time of 45 min with a 0.1 mL sample volume
- Read at 600–660 nm (635 nm optimal) on a microplate reader
- Supplied for 96 Assays with a 96-well microplate and plate sealers
Blood protein is precipitated with a protein precipitator, which destroys enzyme activity and prevents the in vitro formation of free ammonia. Most interfering colored substances are removed at the same time. In the non-protein filtrate, indigo is formed by the Berthelot reaction, and color depth is proportional to blood ammonia content. Blood ammonia content is determined by comparison with a standard curve.
- Measurement of blood ammonia in serum and plasma for liver biomarker research
- Studies of amino acid and ammonia metabolism
- Investigation of protein metabolism and deamination-related ammonia production
- Comparative ammonia measurements across species, including human, mouse, rat, dog, and horse samples
| Component | Quantity | Storage |
| Reagent 1: Acid Reagent | 40 mL × 1 vial | 2–8°C, 12 months |
| Reagent 2: Chromogenic Agent A | 20 mL × 1 vial | 2–8°C, 12 months, protect from light |
| Reagent 3: Chromogenic Agent B | 20 mL × 1 vial | 2–8°C, 12 months, protect from light |
| Reagent 4: 7 mmol/L Standard | 1.5 mL × 1 vial | 2–8°C, 12 months |
| Reagent 5: Standard Diluent | 30 mL × 1 vial | 2–8°C, 12 months |
- Microplate reader capable of 600–660 nm reading (635 nm optimal)
- Micropipettors and disposable tips
- Centrifuge capable of 1100 × g
- Incubator set to 37°C
- Vortex mixer
- 1.5 mL EP tubes
- Double-distilled water or normal saline (0.9% NaCl) for sample dilution if needed
- Phosphate-buffered saline (0.01 M, pH 7.4), optional diluent
Serum/plasma
- Serum and plasma samples are detected directly without further preparation.
- Avoid hemolysis during collection and testing, because red blood cell ammonia content is 2.8 times higher than that of plasma.
- Test samples promptly. Samples can be stored at 2–8°C for 2–4 hours or at −20°C for 24 hours.
- Seal the sample tube immediately after sampling to avoid ammonia spillage.
- Recommended dilution factor is 1 for human serum, human plasma, mouse serum, rat plasma, dog serum, and horse serum.
- Diluent is double distilled water or normal saline (0.9% NaCl).
- For other sample types, perform a pretest to confirm the dilution factor.
Reagent preparation
- Equilibrate all reagents to room temperature before use.
- Prepare a fresh set of standards for each assay and discard working standards after use.
- Prepare standards by mixing 7 mmol/L standard and standard diluent: 0 mmol/L (0 µL standard + 350 µL diluent); 0.2 mmol/L (10 µL + 340 µL); 0.4 mmol/L (20 µL + 330 µL); 0.8 mmol/L (40 µL + 310 µL); 1.0 mmol/L (50 µL + 300 µL); 1.5 mmol/L (75 µL + 275 µL); 2.0 mmol/L (100 µL + 250 µL); 2.5 mmol/L (125 µL + 225 µL).
- Centrifuge small-volume reagents before use.
Sample preparation
- Serum and plasma are tested directly.
- Seal sample tubes immediately after collection to prevent ammonia loss.
- Dilute samples if the expected value is outside the detection range, using double-distilled water or normal saline (dilution factor 1 for the listed sample types).
Assay procedure
- Add 100 µL of each standard to labeled 1.5 mL EP tubes. Add 100 µL of sample to sample tubes.
- Add 300 µL of Acid Reagent, mix fully by vortexing, and centrifuge at 1100 × g for 10 min.
- Complete the chromogenic reaction within 20 min. Transfer 40 µL of clarified supernatant from each tube to the corresponding microplate well.
- Add 120 µL of Chromogenic Agent A, then 120 µL of Chromogenic Agent B, successively. Do not mix A and B before adding.
- Mix on the microplate reader for 5 s, then incubate at 37°C for 25 min.
- Read the OD value of each well at 635 nm.
Calculation
- Average the duplicate readings for each standard.
- Subtract the mean OD of the blank (0 mmol/L standard) from all standard readings to obtain absolute OD values.
- Plot absolute OD (y) against concentration in mmol/L (x) and fit a linear standard curve (y = ax + b).
- Calculate blood ammonia (mmol/L) = (ΔA635 − b) ÷ a × f, where ΔA635 = OD sample − OD blank and f = sample dilution factor.
Blood ammonia (mmol/L) = (ΔA635 - b) ÷ a × f
- ΔA635 is the absorbance at 635 nm of the sample minus the blank (ΔA635 = OD Sample - OD Blank).
- a is the slope and b is the intercept of the standard curve y = ax + b, where y is the absolute OD value and x is the standard concentration in mmol/L.
- Blood ammonia is reported in mmol/L.
- f is the dilution factor of the sample before testing. Use f = 1 for the listed sample types without dilution; multiply by the appropriate factor for diluted samples.
Which sample types can be tested with this kit?
The kit measures blood ammonia in serum and plasma. The recommended dilution factor is 1 for human serum, human plasma, mouse serum, rat plasma, dog serum, and horse serum. For other sample types, perform a pretest to confirm the dilution factor.
How should samples be stored before testing?
Test samples promptly. Samples can be stored at 2–8°C for 2–4 hours or at −20°C for 24 hours. Seal samples immediately after sampling to avoid ammonia spillage.
How should kit components be stored?
Store reagents at 2–8°C for 12 months. Chromogenic Agent A, Chromogenic Agent B, and the reagents should be protected from light. Equilibrate all reagents to room temperature before use.
What should I do if my result is outside the detection range?
The detection range is 0.01–2.5 mmol/L. If the concentration is not within this range, take an extra dilution or concentration of the sample and retest.
What are the key points for a reliable assay?
Hemolysis must be avoided because red blood cells contain 2.8 times more ammonia than plasma. The supernatant after centrifugation must be clear, and the chromogenic reaction must be completed within 20 min. Chromogenic Agents A and B must not be mixed before adding, and disposable materials are recommended to avoid contamination.
What instrument settings and assay time are required?
Read absorbance on a microplate reader at 600–660 nm, with an optimum wavelength of 635 nm. Incubate at 37°C for 25 min after adding the chromogenic agents. The total assay time is 45 min.
| Storage: | 2–8°C, protect from light |
| Shelf Life: | 12 months |
| Shipping: | Gel pack |
Notes
- For Research Use Only. Not for use in clinical diagnosis.
- Avoid hemolysis. Ammonia content of red blood cells is 2.8 times higher than that of plasma.
- Test samples promptly. Glutamine and peptides can hydrolyze and release ammonia. Samples may be stored at 2–8°C for 2–4 hours or at -20°C for 24 hours.
- Seal sample tubes immediately after sampling to avoid ammonia loss.
- Complete the chromogenic reaction within 20 min after the acid reagent step. Clarify the supernatant by centrifugation before transfer.
- Chromogenic Agent A and Chromogenic Agent B must not be mixed before adding.
- Use disposable materials to avoid contamination with interfering substances.
- If the concentration is outside the detection range, dilute or concentrate the sample.
- Run a pre-test for sample types not listed in the protocol.
- Wear a lab coat and gloves during use.