Glucose Uptake Fluorometric Assay Kit
The Glucose Uptake Fluorometric Assay Kit (MAES0012) quantifies glucose uptake in cultured cell samples using the glucose analog 2-deoxyglucose (2-DG). Intracellular 2-DG-6-phosphate is detected through an NADPH-linked enzymatic reaction that generates a fluorescent product read at Ex/Em 530/590 nm. Researchers use the kit to study cell signaling and glucose metabolism by comparing uptake across experimental conditions.
The Glucose Uptake Fluorometric Assay Kit (MAES0012) measures glucose uptake in cell samples. It uses 2-deoxyglucose (2-DG), which cells take up and convert to 2-DG-6P. Glucose uptake content is reported in nmol/µL and is calculated from a standard curve prepared with known concentrations of a 0.3 nmol/µL standard. The readout therefore reflects the uptake and intracellular conversion of 2-DG in the cells being tested.
Researchers use the kit to quantify how much glucose cells take up under defined experimental conditions. The protocol includes a control well and a sample well for each condition, which allows uptake to be compared between untreated and treated cells. The kit is intended for measuring glucose uptake content in cell samples, and the manual's worked example applies the protocol to cultured 293T cells.
After uptake, 2-DG-6P is converted by glucose dehydrogenase to 6PDG, and NADP+ is reduced to NADPH. Diaphorase then uses NADPH to convert a probe into fluorescent substances. In the workflow, cells are incubated with 2-DG at 37°C for 30 min and washed, then treated with acid reagent for 10 min at room temperature and alkali reagent. A 30 µL aliquot from each standard, sample, or control well is transferred to a fluorescence well, and 170 µL of enzyme working solution is added. After a 30 min incubation at 37°C, fluorescence is read at an excitation wavelength of 530 nm and an emission wavelength of 590 nm.
Cells should be seeded at 20000-50000 cells per well in a 96-well plate and cultured or treated according to the experimental design. Before the uptake step, cells are starved overnight in serum-free medium, or for 6 to 8 hours, with attention to cell condition. The medium is then discarded, and cells are washed twice with 200 µL of KRPH solution containing 2% BSA. KRPH solution contains 20 mM Hepes, 5 mM KH2PO4, 1 mM MgSO4, 1 mM CaCl2, 136 mM NaCl, 4.7 mM KCl, 2% BSA, at pH 7.4. Kit reagents should be stored at -20°C and equilibrated to room temperature before use.
The manual gives a detection range of 0.02-0.3 nmol/µL. In the example analysis, the standard curve was y = 48152x + 521.17, and 293T cells (0.7×10^6 cells) gave a glucose uptake content of 0.093 nmol/µL. The manual states that standard curve values are provided for reference only, because fluorescence values vary with operator, pipetting technique, and temperature. Each standard curve should therefore be generated in the same run as the samples.
| Product Name: | Glucose Uptake Fluorometric Assay Kit |
| SKU: | MAES0012 |
| Size: | 96 Assays |
| Detection Method: | Fluorometric |
| Sample Type: | Cells |
| Assay Time: | 100 min (1 h 40 min) |
| Instrument: | Fluorescence microplate reader |
| Excitation/Emission: | 530 nm / 590 nm |
| Detection Range: | 0.02-0.3 nmol/µL |
| Detection Method: | Fluorometric |
| Sample Type: | Cells |
| Detection Range: | 0.02-0.3 nmol/µL |
| Sensitivity (lower limit): | 0.02 nmol/µL |
| Excitation/Emission: | 530 nm / 590 nm |
| Assay Time: | 100 min |
- Fluorometric readout at Ex/Em 530/590 nm on a fluorescence microplate reader
- Detection range 0.02-0.3 nmol/µL with a 0.02 nmol/µL lower limit
- Uses 2-DG, a glucose analog that accumulates intracellularly as 2-DG-6-phosphate
- Approximately 100 min total assay time
- Includes 2-DG, standard, enzyme reagent powder, and buffer components for 96 assays
Intracellular 2-DG is converted to 2-DG-6P, which is catalyzed by glucose dehydrogenase to produce 6PDG, while NADP+ is converted to NADPH. The generated NADPH converts the probe into fluorescent substances under the action of diaphorase. Glucose uptake is calculated from fluorescence intensity measured at an excitation wavelength of 530 nm and an emission wavelength of 590 nm.
- Measuring glucose uptake in cultured cell samples
- Studying glucose metabolism and its regulation in cell models
- Investigating cell signaling pathways that influence glucose transport
- Comparing 2-DG uptake across treatment or genotype conditions in a 96-well format
| Component | Quantity | Storage |
| Acid Reagent | 10 mL × 1 vial | -20°C |
| Alkali Reagent | 10 mL × 1 vial | -20°C |
| Chromogenic Agent | 25 mL × 1 vial | -20°C, protect from light |
| Enzyme Reagent (powder) | 2 vials | -20°C |
| 10 mmol/L 2-DG | 1.5 mL × 1 vial | -20°C |
| 0.3 nmol/µL Standard | 2 mL × 1 vial | -20°C |
| KRPH Buffer Solution | 55 mL × 1 vial | -20°C |
| Black 96-well microplate | 1 plate (96 wells) | No requirement |
| Plate sealer | 2 pieces | No requirement |
| Sample layout sheet | 1 sheet | No requirement |
- Fluorescence microplate reader (Ex/Em 530 nm/590 nm)
- Micropipettor and tips
- Incubator set to 37°C
- Water bath
- Double distilled water (for standard dilutions)
- Serum-free cell culture medium
- Cultured cells (seeding density 20,000-50,000 cells per well)
- Centrifuge (for brief spin of small reagent volumes)
Cells
- Seed cells at a density of 20000-50000 cells per well in a 96-well plate. Culture and treat cells according to the experimental needs.
- Starve cells overnight in serum-free cell medium. Starvation can be 6 to 8 hours; monitor the state of the cells.
- Discard the medium and wash cells twice with 200 µL of KRPH solution (including 2% BSA).
- Add 100 µL of KRPH solution (including 2% BSA) to the control well and sample well.
- Add 10 µL of 10 mmol/L 2-DG to the sample well. Add 10 µL of KRPH solution to the control well.
- Incubate at 37°C for 30 min.
- Wash cells 3 times with 100 µL of KRPH solution.
- Add 50 µL of acid reagent and stand at room temperature for 10 min, then add 50 µL of alkali reagent.
- Transfer 30 µL of each standard, sample well, or control well into the corresponding fluorescence wells.
- Add 170 µL of enzyme reagent working solution to each well.
- Incubate at 37°C for 30 min, then measure fluorescence.
- If the sample concentration is not within the detection range, take an extra dilution or concentration of the sample. It is recommended to run a pre-test if the sample type is not listed in the manual.
Reagent preparation
- Equilibrate all reagents to room temperature before use.
- Prepare enzyme working solution by dissolving one vial of enzyme reagent powder in 10 mL of chromogenic agent and mixing well. Store at 2-8°C for up to 3 days, protected from light.
- Prepare fresh standards each time by diluting the 0.3 nmol/µL standard with double distilled water. Recommended gradient: 0, 0.06, 0.12, 0.15, 0.18, 0.21, 0.24, and 0.3 nmol/µL (for example, 40 µL of standard plus 160 µL of water for 0.06 nmol/µL in a 200 µL total volume). Discard working dilutions after use.
Assay steps
- Seed cells at 20,000-50,000 cells per well in a 96-well plate and culture or treat as the experiment requires.
- Starve cells in serum-free medium (overnight per the manual; the manual notes 6-8 h is also acceptable, monitoring cell condition). Discard medium and wash cells twice with 200 µL KRPH solution containing 2% BSA.
- Add 100 µL KRPH solution (with 2% BSA) to control and sample wells. Add 10 µL of 10 mmol/L 2-DG to sample wells and 10 µL of KRPH solution to control wells. Incubate at 37°C for 30 min.
- Wash cells 3 times with 100 µL KRPH solution. Add 50 µL acid reagent and stand at room temperature for 10 min, then add 50 µL alkali reagent.
- Transfer 30 µL of each standard, sample, or control into the corresponding wells of a fresh black fluorescence plate.
- Add 170 µL enzyme reagent working solution to each well.
- Incubate at 37°C for 30 min.
- Read fluorescence at Ex 530 nm / Em 590 nm.
Glucose uptake content (nmol/µL) = (F2 - F1 - b) ÷ a
- F2 is the fluorescence intensity of the sample well.
- F1 is the fluorescence intensity of the control well.
- a is the slope and b is the intercept of the standard curve y = ax + b, generated from absolute fluorescence values (average standard reading minus the blank, Standard #1) plotted against standard concentration in nmol/µL.
- Standard readings are averaged in duplicate before the blank is subtracted.
- The result is expressed in nmol/µL. The formula contains no dilution factor term; diluted or concentrated samples should be handled as described in the sample preparation steps.
Which sample types are suitable for this kit?
The kit is intended for measuring glucose uptake content in cell samples. Cells should be seeded at 20000-50000 cells per well in a 96-well plate, and the manual's example uses 293T cells.
How should the kit components be stored?
Most components should be stored at -20°C for 12 months, and the chromogenic agent should be kept protected from light. After reconstitution, the enzyme reagent working solution should be stored at 2-8°C for 3 days, protected from light. Working standard dilutions must be discarded after use, and reagents from different kits must not be mixed.
What should I do if my sample is outside the detection range?
The detection range is 0.02-0.3 nmol/µL. If the concentration is not within this range, take an extra dilution or concentration of the sample before the assay.
What settings are needed for the instrument?
Use a fluorescence microplate reader with excitation at 530 nm and emission at 590 nm. The protocol uses a black microplate.
How long does the assay take?
The assay time is 1 h 40 min.
| Storage: | -20°C, protect from light |
| Shelf Life: | 12 months |
| Shipping: | Ice bag |
Notes
- For Research Use Only. Not for use in clinical diagnosis or therapeutic procedures.
- Protect the chromogenic agent and the enzyme working solution from light.
- Prepare standard dilutions fresh and discard after use.
- Do not mix reagents from different kit lots.
- Centrifuge small reagent volumes briefly before use.
- If sample values fall outside the detection range, dilute or concentrate the sample and re-assay.
- Run a pre-test if the sample type is not listed in the manual.
- Wear a lab coat and gloves during handling.