Description
Urine Protein Colorimetric Assay Kit
The Urine Protein Colorimetric Assay Kit (MAES0259) quantifies total protein in urine samples by a Coomassie Blue G-250 dye-binding reaction read on a microplate reader. The assay runs in a 96-well format using 30 µL of standard or sample and a single chromogenic reagent addition. It is intended for research use only.
Urinary protein is the total protein filtered from plasma by the glomerulus and excreted in urine. In healthy kidneys, most of this filtered protein is reabsorbed by the renal tubules, so only small amounts appear in the final urine. Protein concentration in urine therefore reflects how effectively the kidney system retains protein. When reabsorption capacity declines, urinary protein rises, which is why this measurement serves as an auxiliary basis for judging kidney disease.
Researchers measure urine protein to study kidney physiology and renal function, a research area this kit is positioned within. Elevated urinary protein is a recognized feature of kidney disease, so quantifying it supports studies that examine renal damage, filtration, and tubular reabsorption. The kit is intended for research use only and is not validated for clinical diagnosis or other clinical decisions.
The assay uses Coomassie Blue G-250 as the colorimetric reagent. Under acidic conditions, protein in the sample binds the dye and the reaction shifts from brownish red to blue. The blue product has a maximum absorption between 585 and 605 nm, and readings are taken at 595 nm, the recommended optimum wavelength. Each well receives 30 µL of standard or sample followed by 180 µL of chromogenic agent. After a brief 5 s mix on the plate reader and a 5 min room temperature incubation, optical density (OD) is read on a microplate reader. The total assay time is about 30 min.
Fresh urine is the validated sample type. Samples should be centrifuged at 10000×g for 15 min at 4°C, and the clear supernatant should be kept on ice until use. Standards are prepared from a 500 mg/L stock by serial dilution in double distilled water over 0 to 100 mg/L, and a fresh standard set should be prepared for each run. Other sample types are not covered by the validated protocol and require a pretest to confirm suitable dilution.
The kit reports a sensitivity of 7.77 mg/L and a detection range of 7.77 to 100 mg/L. Intra-assay precision is 2.4% CV and inter-assay precision is 4.7% CV. Recovery averaged 105% across high, middle, and low concentration samples tested six times each. Samples that fall outside the detection range require additional dilution or concentration before testing.
| Product Name: | Urine Protein Colorimetric Assay Kit |
| SKU: | MAES0259 |
| Sample Type: | Urine |
| Assay Type: | Quantitative |
| Detection Method: | Colorimetric |
| Detection Instrument: | Microplate reader (585-605 nm, optimum 595 nm) |
| Research Area: | Liver and renal function |
| Sensitivity: | 7.77 mg/L |
| Detection Range: | 7.77-100 mg/L |
| Intra-assay CV: | 2.4% |
| Inter-assay CV: | 4.7% |
| Sample Volume: | 30 µL |
| Valid Period: | 12 months |
- Quantitative colorimetric measurement of urine protein
- Coomassie Blue G-250 chemistry; color shifts from brownish red to blue
- Detection range 7.77-100 mg/L with 7.77 mg/L analytical sensitivity
- Read at 585-605 nm; optimum wavelength 595 nm
- Intra-assay CV 2.4% and inter-assay CV 4.7%
- Stable at -20°C for 12 months, protected from light
The protein in urine reacts with Coomassie Blue G-250 under acidic conditions to produce a chromogenic substance. The reaction system changes from brownish red to blue. The blue substance has a maximum absorption wavelength at 585-605 nm, and absorbance is read at 595 nm. Absorbance is blank-corrected and compared against a linear standard curve to calculate urine protein concentration.
- Quantification of total urine protein in kidney function research
- Comparison of urine protein levels across experimental or treatment groups
- Liver and renal function studies requiring a rapid colorimetric protein readout
- Screening of urine samples for protein content before further analysis
| Component | Quantity | Storage |
| Chromogenic Agent | 1 vial | -20°C, 12 months, protect from light |
| Reagent 2: 500 mg/L Standard Solution | 1 vial | -20°C, 12 months, protect from light |
| Microplate | 1 microplate | No requirement |
| Plate Sealer | 1 plate sealer | — |
| Sample Layout Sheet | 1 sheet | — |
- Microplate reader capable of 585-605 nm reading (optimum 595 nm)
- Centrifuge capable of 10000×g at 4°C
- Double-distilled water (diluent for standards and urine samples)
- Pipettes and tips for 30 µL, 60-200 µL volumes
- Microcentrifuge tubes
- Graphing software or Excel for standard curve fitting
Urine
- Collect fresh urine.
- Centrifuge the urine at 10000×g for 15 min at 4°C.
- Collect the supernatant and keep it on ice until detection.
- Dilution: the recommended dilution factor for human urine is 1, so no dilution is required before testing.
- If the urine protein concentration is outside the 7.77 to 100 mg/L detection range, perform an extra dilution or concentration of the sample and account for it in the calculation.
Sample preparation
- Collect fresh urine and centrifuge at 10000×g for 15 min at 4°C.
- Collect the supernatant and keep on ice until assay.
- Use human urine at a dilution factor of 1.
Standard preparation
- Equilibrate all reagents to room temperature before use.
- Dilute 200 µL of 500 mg/L standard solution with 800 µL double-distilled water to make 100 mg/L standard.
- Prepare the standard series fresh by diluting the 100 mg/L standard with double-distilled water to 0, 30, 40, 60, 70, 80, 90, and 100 mg/L (for example, 60 µL of 100 mg/L standard plus 140 µL water for 30 mg/L). Discard working dilutions after use.
Assay
- Add 30 µL of each standard to standard wells and 30 µL of sample to sample wells.
- Add 180 µL of chromogenic agent to each well.
- Mix on the plate reader for 5 s, then stand at room temperature for 5 min.
- Measure OD at 595 nm with the microplate reader.
Calculation
- Average the duplicate readings for each standard.
- Subtract the mean OD of the blank (standard 0 mg/L) from all standard readings to obtain absolute OD.
- Plot absolute OD (y) against concentration (x) and fit a linear equation (y = ax + b).
- Calculate urine protein (mg/L) = (ΔA595 − b) ÷ a × f, where ΔA595 = OD sample − OD blank and f is the sample dilution factor.
urine protein content (mg/L) = (ΔA595 − b) ÷ a × f
- ΔA595 is the sample OD at 595 nm minus the blank OD at 595 nm (ΔA595 = OD Sample − OD Blank).
- a is the slope and b is the intercept of the standard curve, y = ax + b, where y is the absolute OD (standard OD minus the mean blank OD) and x is the standard concentration in mg/L.
- f is the dilution factor applied to the sample before testing. For human urine tested without dilution, f = 1.
- The result is reported in mg/L.
- Worked example from the manual: with standard curve y = 0.0019x − 0.0056, control OD 0.293, and sample OD 0.399 for 30 µL of human urine, the result is (0.399 − 0.293 + 0.0056) ÷ 0.0019 = 58.73 mg/L.
What sample type is validated for this kit?
The kit is validated for urine. For human urine, the recommended dilution factor is 1. Other sample types require a pretest to confirm the dilution factor.
How should urine samples be handled before testing?
Centrifuge fresh urine at 10000×g for 15 min at 4°C. Collect the supernatant and keep it on ice until detection.
How should the kit be stored, and how long is it valid?
Store the kit at -20°C, protected from light. The valid period is 12 months.
What is the detection range, and what should I do if my sample is outside it?
The detection range is 7.77 to 100 mg/L. If a sample falls outside this range, perform an extra dilution or concentration before testing.
What instrument settings should I use?
Read optical density on a microplate reader at 595 nm, within the 585 to 605 nm range. Use 30 µL of sample or standard and 180 µL of chromogenic agent per well.
How long does the assay take, and what affects results?
The assay takes about 30 min, including a 5 min room temperature incubation after mixing. Results are closely related to reagent condition, operation, and environment, so follow the protocol strictly.
| Storage Temperature: | -20°C |
| Valid Period: | 12 months |
| Light Protection: | Store in the dark (shading light) |
| Microplate: | No storage requirement |
Notes
- For Research Use Only. Not for use in clinical diagnosis or therapeutic procedures.
- Read the instructions completely and adjust the instrument before running the assay.
- Wear a lab coat and latex gloves during use.
- Dilute or concentrate samples that fall outside the 7.77-100 mg/L detection range.
- Run a pretest for any sample type not listed in the instructions.
- Do not mix components between kits.
- Centrifuge small reagent volumes before use to ensure sufficient recovery.
- Results are sensitive to reagent handling, operator technique, and temperature; include a fresh standard curve for each run.