Description
H2S Colorimetric Assay Kit
The H2S Colorimetric Assay Kit (MAES0185) quantifies hydrogen sulfide (H2S) in serum, plasma, and animal tissue samples using an indirect methylene blue method. H2S reacts in the presence of Fe3+ to form methylene blue, which is read at 665 nm on a microplate reader. Researchers use the kit to study H2S signaling and to measure H2S in rodent models, including studies of H2S donors.
Hydrogen sulfide (H2S) is the analyte measured by this kit, which quantifies H2S content in serum, plasma, and animal tissue samples. The cited literature connects H2S to mitochondrial biology, including a study of sulfide quinone oxidoreductase in voltage sensing of the mitochondrial permeability transition pore and work on mitochondria-targeted H2S donors. The kit is intended for research use only and is not for clinical diagnosis.
Researchers use H2S measurements in the physiological and disease contexts represented by the cited publications. These include cancer imaging and therapy with H2S-responsive polymers and nanoplatforms for photodynamic therapy, fatty liver and obesity in mice fed a high-fat diet, inflammation during infected wound healing, and the progression of adenomatous polyps. Measurements in these studies rely on a consistent, quantitative readout of H2S in biological matrices.
The method detects H2S indirectly. H2S reacts with acetate solution to form ZnS, which is dissolved in alkaline solution. In the presence of Fe3+ and the chromogenic agent, stable methylene blue forms, with a maximum absorption peak at 665 nm. Sample and control tubes are processed in parallel, with double distilled water replacing sample in the control. After precipitation, washing, and reagent additions, 250 µL of supernatant is transferred to microplate wells, incubated for 10 min at room temperature, and read at 665 nm.
Serum and plasma are tested directly. Tissue is homogenized in cold normal saline, and the supernatant is used after centrifugation. Serum and plasma that cannot be tested on the same day can be stored at -80°C for one month. Kit components are stored at 2-8°C for 12 months. The chromogenic agent and ferric salt reagent must be protected from light, and all reagents must be equilibrated to room temperature before use.
The assay reports a sensitivity of 0.15 µmol/L, determined by adding two standard deviations to the mean OD of a zero standard assayed 20 times. The detection range is 0.76-100 µmol/L. Intra-assay precision for three human serum samples ranged from 2.6% to 3.5% CV across 20 replicates. Inter-assay precision ranged from 9.5% to 10.4% CV across 17 runs in duplicate by three operators. Recovery averaged 94%, with individual recoveries of 89% to 99% at low, middle, and high concentrations.
| Product Name: | H2S Colorimetric Assay Kit |
| SKU: | MAES0185 |
| Size: | 96 Assays |
| Detection Method: | Colorimetric (indirect methylene blue) |
| Sample Type: | Serum, plasma, animal tissue |
| Detection Wavelength: | 665 nm (microplate reader, 660-670 nm filter range) |
| Instrument: | Microplate reader |
| Assay Time: | 1 h 20 min |
| Research Area: | Other |
| Use: | Research Use Only |
| Sensitivity: | 0.15 µmol/L |
| Detection Range: | 0.76-100 µmol/L |
| Recovery Rate: | 94% (average) |
| Intra-assay CV: | 3.1% |
| Inter-assay CV: | 9.9% |
| Sample Volume: | 100 µL |
| Assay Time: | 1 h 20 min |
| Measuring Wavelength: | 665 nm |
- Indirect methylene blue colorimetric method, read at 665 nm
- Validated for serum, plasma, and animal tissue homogenate samples
- Sensitivity of 0.15 µmol/L; detection range 0.76-100 µmol/L
- Intra-assay CV 3.1% and inter-assay CV 9.9%
- Total assay time of 1 h 20 min; 100 µL sample volume
- Supplied as a 96 Assay kit with microplate and plate sealer
In the presence of Fe3+, H2S is converted to methylene blue, which has a maximum absorption peak at 665 nm. H2S reacts with the acetate solution to form ZnS, which dissolves in alkaline solution. The H2S content is calculated indirectly from the OD at 665 nm, using the methylene blue molar extinction coefficient (9.4 × 10-3 L/µmol/cm) and an optical path of 0.7 cm. Serum and plasma results are reported in µmol/L; tissue results are normalized to protein concentration (µmol/gprot).
- Quantifying H2S in serum and plasma from rodent studies
- Measuring H2S in animal tissue homogenates, such as brain, kidney, and spleen
- Investigating H2S roles in nervous and circulatory system signaling
- Evaluating H2S donor compounds in animal disease models, such as high-fat diet and cancer models
| Component | Quantity | Storage |
| Reagent 1: Buffer Solution | 25 mL × 1 vial | 2-8°C, 12 months |
| Reagent 2: Alkali Reagent | 12 mL × 1 vial | 2-8°C, 12 months |
| Reagent 3: Chromogenic Agent | 12 mL × 1 vial | 2-8°C, 12 months, protect from light |
| Reagent 4: Protein Precipitator | 12 mL × 1 vial | 2-8°C, 12 months |
| Reagent 5: Ferric Salt Reagent | 2.5 mL × 1 vial | 2-8°C, 12 months, protect from light |
| Microplate | 96 wells | No special storage requirement |
| Plate Sealer | 2 pieces | Not applicable |
| Sample Layout Sheet | 1 piece | Not applicable |
- Microplate reader capable of 660-670 nm measurement
- Micropipettor
- Centrifuge capable of 12,000 × g at 4°C
- Incubator
- Vortex mixer
- Double-distilled water
- Normal saline (0.9% NaCl)
- PBS (0.01 M, pH 7.4) for tissue washing and dilution of other sample types
- Homogenizer (Dounce type) for tissue samples
- Protein concentration assay for tissue supernatant
Serum/plasma
- Detect serum and plasma directly.
- If samples are not detected on the same day, store serum or plasma at -80°C for up to one month.
- Dilution factor: human plasma 1; mouse plasma 1. Dilute with normal saline (0.9% NaCl).
Tissue
- Harvest the amount of tissue needed for each assay (initial recommendation 20 mg).
- Wash tissue in cold PBS (0.01 M, pH 7.4).
- Homogenize 20 mg tissue in 180 µL normal saline (0.9% NaCl) with a dounce homogenizer at 4°C.
- Centrifuge at 10000×g for 10 min to remove insoluble material. Collect the supernatant and keep it on ice for detection.
- Determine the protein concentration of the supernatant separately, as required for calculation.
- Dilution factor: 10% rat spleen, kidney, and brain tissue homogenates each have a dilution factor of 1. Dilute with normal saline (0.9% NaCl). For other sample types, run a pretest to confirm the dilution factor.
- Do not take the sediment when adding the supernatant to the microplate.
Sample preparation
- Serum and plasma: assay directly. If not assayed the same day, store at -80°C for up to one month.
- Tissue: wash about 20 mg tissue in cold PBS (0.01 M, pH 7.4), then homogenize in 180 µL normal saline (0.9% NaCl) at 4°C.
- Centrifuge the homogenate at 10,000 × g for 10 min. Collect the supernatant, keep it on ice, and determine its protein concentration.
- Dilute samples as needed. Recommended dilution factors are 1 for human and mouse plasma and 1 for 10% rat spleen, kidney, and brain tissue homogenates. Use normal saline (0.9% NaCl) as diluent; pretest other sample types.
Assay procedure
- Equilibrate all reagents to room temperature before use.
- Add 100 µL Reagent 1 (buffer) to the control tube and to the sample tube.
- Add 100 µL double-distilled water to the control tube and 100 µL sample to the sample tube.
- Add 100 µL Reagent 2 (alkali reagent) to each tube and vortex for 3 s.
- Centrifuge at 12,000 × g for 10 min at 4°C. Discard the supernatant and keep the pellet.
- Add 150 µL double-distilled water and vortex for 3 s.
- Centrifuge at 12,000 × g for 10 min at 4°C. Discard the supernatant and keep the pellet.
- Add 100 µL Reagent 1 (buffer) to each tube.
- Add 100 µL Reagent 3 (chromogenic agent) to each tube and vortex for 10 s.
- Add 20 µL Reagent 5 (ferric salt reagent) to each tube. Stand for 30 s, then vortex for 5 s.
- Add 100 µL Reagent 4 (protein precipitator) to each tube and vortex for 3 s.
- Centrifuge at 12,000 × g for 10 min at 4°C. Transfer 250 µL of supernatant to the corresponding microplate wells. Do not transfer the sediment.
- Stand for 10 min at room temperature, then read OD at 665 nm on a microplate reader.
Serum (plasma) sample: H2S (µmol/L) = ΔA665 / (ε × b) × f Tissue sample: H2S (µmol/gprot) = ΔA665 / (ε × b) × f ÷ Cpr
- ΔA665 = OD(sample) − OD(control), measured at 665 nm (dimensionless).
- ε = molar extinction coefficient of methylene blue at 665 nm, 9.4×10-3 L/µmol/cm.
- b = optical path length, 0.7 cm.
- f = dilution factor of the sample before testing. Multiply by f for diluted samples (use f = 1 when no dilution is applied).
- Cpr = protein concentration of the sample in gprot/L. Divide by Cpr to express tissue results per gram of protein (µmol/gprot).
Which sample types can be tested with this kit?
The kit is intended for serum, plasma, and animal tissue samples. Human plasma, mouse plasma, and 10% rat spleen, kidney, and brain tissue homogenates are listed with a dilution factor of 1. Other sample types require a pretest to confirm the dilution factor.
How should samples and kit components be stored?
Serum and plasma that are not tested on the same day can be stored at -80°C for one month. Kit components are stored at 2-8°C for 12 months, and the chromogenic agent and ferric salt reagent must be protected from light.
What is the detection range, and what should I do if my sample is out of range?
The detection range is 0.76-100 µmol/L. If the H2S concentration is not within this range, perform an extra dilution or concentration of the sample.
What should I watch for during the assay?
Do not take the sediment when adding the supernatant to the microplate, as this affects the result. Do not mix reagents from different kits, and centrifuge small reagent volumes before use. Results are sensitive to reagents, operation, and environment, so follow the steps strictly.
What instrument settings and assay time apply?
Read the OD value at 665 nm on a microplate reader with a 660-670 nm filter after standing for 10 min at room temperature. The total assay time is 1 h 20 min.
| Storage: | 2-8°C |
| Shelf Life: | 12 months |
| Light Sensitivity: | Chromogenic Agent and Ferric Salt Reagent: protect from light (shading light) |
| Shipping: | Gel pack |
Notes
- For Research Use Only. Not for use in clinical diagnosis or other clinical purposes.
- Store reagents strictly at 2-8°C. Do not mix reagents from different kits.
- Centrifuge vials containing small volumes before use to recover sufficient reagent.
- Do not take the sediment when transferring supernatant to the microplate.
- Equilibrate all reagents to room temperature before use.
- If a sample is outside the detection range, dilute or concentrate the sample and repeat the assay.
- Take a pretest for any sample type not listed in the manual.
- Wear a lab coat and gloves during the assay.