One-step TUNEL Flow Cytometry Apoptosis Kit (Red, AF555)
The One-step TUNEL Flow Cytometry Apoptosis Kit (Red, AF555) labels the exposed 3'-OH ends of fragmented genomic DNA with fluorescein-labeled dUTP using Terminal Deoxynucleotidyl Transferase (TdT). Labeled apoptotic cells are quantified by flow cytometry. Researchers use the kit to detect apoptosis in suspension and adherent cell samples with a fast, simple workflow of about 2.5-3 hours.
Apoptosis is programmed cell death. During apoptosis, specific DNA endonucleases are activated and cut genomic DNA between nucleosomes. The DNA of apoptotic cells is cleaved into multimers of 180~200 bp fragments, corresponding to the oligonucleosomal size, and typically migrates as a ladder of 180~200 bp on an agarose gel. The breaks expose 3'-OH ends on the fragmented DNA, and these free 3'-OH ends are the substrate this kit detects.
Researchers use this kit to detect apoptosis in both suspension and adherent cell populations by flow cytometry. Because the readout is per cell, the kit supports quantification of the apoptotic fraction within a population rather than only a bulk DNA ladder. The manual recommends setting up a negative control in each experiment to show the objectivity and accuracy of the TUNEL reaction. Where apoptosis is not induced, the manual advises testing different concentration gradients and induction times to find suitable induction conditions. The validation data use HL-60 cells treated with camptothecin as a positive apoptosis model.
The assay uses a one-step TUNEL labeling reaction. Terdeoxynucleotidyl Transferase (TdT) catalyzes the addition of a fluorescent dUTP label to the exposed 3'-OH ends. The label is AF555 (excitation/emission 555/565 nm), and it is read on the PE/PE-TR channel of a flow cytometer. Cells are fixed, permeabilized, equilibrated, and then incubated with the TdT labeling working solution at 37 °C for 60 min in the dark. Following a stop reaction and washes, cells are resuspended in PBS with 1% BSA and analyzed. The total assay time is 2.5 to 3 hours.
The kit is for research use only. It is validated for cells, including suspension cells and adherent cells, and the manual requires a minimum of 5×10^5 cells per sample. Samples should be processed promptly because the fluorescence is easily quenched. For adherent cultures, some apoptotic cells detach into the supernatant, so the supernatant must be collected together with the adherent cells. Using PBS with 1% BSA during resuspension and washing reduces cell loss during centrifugation. Permeabilization must be kept short and cold to avoid cell breakage.
| Product name: | One-step TUNEL Flow Cytometry Apoptosis Kit (Red, AF555) |
| SKU: | AKES114 |
| Detection method: | Fluorometric, flow cytometry |
| Sample type: | Cells (suspension and adherent) |
| Assay time: | 2.5-3 hours |
| Detection instrument: | Flow cytometer |
| Dye: | AF555 |
| Ex/Em: | 555/565 nm |
| Channel: | PE/PE-TR |
| Research use: | Research Use Only |
- TUNEL labeling of DNA strand breaks (exposed 3'-OH ends) detected by flow cytometry
- Red AF555 dye, Ex/Em 555/565 nm, read on PE/PE-TR channel
- Supports suspension and adherent cell samples
- Assay time of 2.5-3 hours
- Requires a minimum of 5 × 10^5 cells per sample
When cells undergo apoptosis, DNA is cut between nucleosomes, exposing 3'-OH ends. Following fixation and permeabilization, TdT catalyzes incorporation of fluorescein-labeled dUTP onto these 3'-OH ends. Labeled cells are detected by flow cytometry in the PE/PE-TR channel, with excitation and emission at 555/565 nm for the AF555 dye.
- Quantifying apoptotic cells in suspension cell cultures by flow cytometry
- Detecting apoptosis in adherent cell samples, including cells detached into the supernatant
- Assessing apoptosis induction in drug-treated or stimulated cell populations
- Apoptosis studies that require DNA fragmentation as a readout alongside other flow cytometry markers
| Component | Storage |
| TdT Equilibration Buffer | -20°C |
| TdT Enzyme | -20°C |
| Labeling Solution (AF555) | -20°C, protect from light |
| Fixation Buffer | -20°C |
| Permeabilization Buffer | -20°C |
| Stop Solution | -20°C |
- PBS buffer with 1% BSA, pH 7.2-7.4
- Centrifuge capable of 300 × g and 600 × g
- Flow cytometer with a PE/PE-TR detection channel
- Water bath or incubator at 37°C
- Ice
- Pipettes and microcentrifuge tubes
Cells (suspension and adherent)
- Collect the cells and count them. Take 1×10^6 cells, centrifuge at 300×g for 5 min, and discard the supernatant. The minimum cell number for this kit is 5×10^5 cells.
- For adherent cells, collect the supernatant as well, because some apoptotic cells are suspended in it. Collect these cells together with the adherent cells.
- Add 200 µL PBS (with 1% BSA) to resuspend the cells, add 200 µL Fixation Buffer, and incubate at room temperature for 60 min. Mix every 15 min.
- Add 1 mL PBS (with 1% BSA), centrifuge at 600×g for 5 min, and discard the supernatant.
- Add 200 µL Permeabilization Buffer and incubate on ice for 10 min. Thaw the Permeabilization Buffer and place it at 4 °C before use. The maximum permeabilization time is 15 min, and a higher temperature or longer time breaks the cells.
- Add 1~2 mL PBS (with 1% BSA) to stop the reaction, mix by gently pipetting, centrifuge at 600×g for 5 min, and discard the supernatant.
- Add 200 µL TdT Equilibration Buffer, mix by gently pipetting, and incubate at 37 °C for 15 min.
- Centrifuge at 600×g for 5 min, discard the supernatant, and collect the cell pellet.
- Prepare the Labeling Working Solution fresh before use. Each sample requires 100 µL: 70 µL TdT Equilibration Buffer, 20 µL Labeling Solution, and 10 µL TdT Enzyme. The negative control uses 80 µL TdT Equilibration Buffer, 20 µL Labeling Solution, and no TdT Enzyme. Gently pipette the mixture; do not vortex.
- Incubate at 37 °C for 60 min in the dark, mixing gently every 20 min.
- Add 1 mL Stop Solution, mix gently by pipetting, and incubate at room temperature for 5 min. Add 300~500 µL PBS (with 1% BSA), centrifuge at 600×g for 5 min, and discard the supernatant.
- Add 200 µL PBS (with 1% BSA) to resuspend the cells, pipetting gently 10~20 times. Analyze by flow cytometry as soon as possible.
Sample preparation and fixation
- Collect cells and count them. Use at least 5 × 10^5 cells per sample. Centrifuge at 300 × g for 5 min and discard the supernatant.
- For adherent cells, collect the floating cells in the supernatant along with the detached cells.
- Resuspend cells in 200 µL PBS (with 1% BSA), add 200 µL Fixation Buffer, and incubate at room temperature for 60 min. Mix every 15 min.
- Add 1 mL PBS (with 1% BSA), centrifuge at 600 × g for 5 min, and discard the supernatant.
Permeabilization and TdT labeling
- Add 200 µL Permeabilization Buffer and incubate on ice for 10 min. Do not exceed 15 min.
- Add 1-2 mL PBS (with 1% BSA) to stop the reaction, mix gently, centrifuge at 600 × g for 5 min, and discard the supernatant.
- Add 200 µL TdT Equilibration Buffer, mix gently, and incubate at 37°C for 15 min.
- Centrifuge at 600 × g for 5 min, discard the supernatant, and keep the cell pellet.
- Prepare Labeling Working Solution fresh before use (100 µL per sample). Gently pipette to mix. Do not vortex.
- Incubate at 37°C for 60 min in the dark, mixing gently every 20 min.
Stop, wash, and analyze
- Add 1 mL Stop Solution, mix gently, and incubate at room temperature for 5 min.
- Add 300-500 µL PBS (with 1% BSA), centrifuge at 600 × g for 5 min, and discard the supernatant.
- Resuspend cells in 200 µL PBS (with 1% BSA) and analyze by flow cytometry in the PE/PE-TR channel.
Which sample types and how many cells are needed?
The kit is used with suspension cells and adherent cells. The minimum number of cells is 5×10^5 per assay, and the manual starts from 1×10^6 cells per sample.
How should the kit and its components be stored?
The kit can be stored at -20 °C for 12 months, protected from light. The TdT Enzyme must be kept strictly at -20 °C and returned to storage immediately after use. Aliquot the Labeling Solution and Fixation and Permeabilization Buffers to avoid freeze-thaw cycles.
What should I do if the apoptotic population is not detected?
Set up different concentration gradients and induction times to find suitable induction conditions. If the sample has too many cells, adjust the cell number or increase the permeabilization. Insufficient fixation or excessive permeabilization time causes cell breakage and cellular debris, so fix cells fully and keep permeabilization to 10~15 min on ice.
Which flow cytometer channel and settings should I use?
Use the PE/PE-TR channel with excitation and emission at 555/565 nm. Use a flow cytometer for analysis and a centrifuge for the wash steps. Include a negative control prepared without TdT Enzyme in every experiment.
What can interfere with the assay?
Buffers containing chelating agents such as EDTA make the TdT enzyme ineffective, so wash cells with a buffer that does not contain a chelating agent. Do not vortex the Labeling Solution or TdT Enzyme. Detect samples as soon as possible to avoid fluorescence quenching.
How long does the assay take?
The assay takes 2.5 to 3 hours from cell collection to flow cytometric analysis.
| Storage temperature: | -20°C |
| Shelf life: | 12 months |
| Light: | Store protected from light |
| Shipping: | Ice bag |
Cautions
- For research use only. Not for use in diagnostic or therapeutic procedures.
- Use a minimum of 5 × 10^5 cells per sample.
- Do not vortex the Labeling Solution or TdT Enzyme.
- Avoid freeze-thaw cycles. Aliquot reagents into smaller volumes on first use.
- Keep the TdT Enzyme strictly at -20°C. Return it to -20°C immediately after use.
- Do not use washing buffers containing chelating agents such as EDTA, which inactivate TdT.
- Keep permeabilization on ice and limit permeabilization to 15 min maximum to avoid cell breakage.
- Detect samples as soon as possible to avoid fluorescence quenching.
- When resuspending cells, pipette gently 10-20 times to avoid cell damage and bubbles.
- Include a negative control (labeling without TdT enzyme) in each experiment.