Description
One-step TUNEL Flow Cytometry Apoptosis Kit (Red, AF647)
The One-step TUNEL Flow Cytometry Apoptosis Kit (Red, AF647) (AKES113) detects apoptotic cells by labeling the exposed 3'-OH ends of fragmented genomic DNA with fluorescein-labeled dUTP using Terminal Deoxynucleotidyl Transferase (TdT). Labeled cells are quantified by flow cytometry in the APC channel. Researchers use the kit to assess apoptosis in suspension and adherent cell populations without gel-based DNA laddering.
Apoptosis is a programmed cell death process in which specific DNA endonucleases are activated and cut genomic DNA between nucleosomes. The resulting fragments are multimers of 180 to 200 bp, corresponding to the oligonucleosomal size, and typically migrate as a ladder of 180 to 200 bp on an agarose gel. The exposed 3'-OH ends generated by this cleavage are the analyte this kit detects, which allows apoptotic cells to be identified by flow cytometry.
Researchers use this kit to detect apoptosis in both suspension and adherent cell populations by flow cytometry. The manual describes the approach as a highly sensitive, fast, and simple method. Its example test compares HL-60 cells treated with 2.5 µM camptothecin for 4 hours against untreated HL-60 cells, which serve as the control. Because the labeling step is performed on fixed and permeabilized cells, the kit is suited to comparing apoptotic cell fractions across treatment conditions.
Terminal deoxynucleotidyl transferase (TdT) catalyzes addition of a labeled dUTP to exposed 3'-OH ends. In this kit the dUTP is labeled with AF647, with excitation and emission at 650 and 670 nm, and the signal is read in the APC channel. Cells are fixed in Fixation Buffer for 60 minutes at room temperature and then permeabilized on ice for up to 15 minutes. The TdT labeling reaction, prepared from TdT Equilibration Buffer, TdT Enzyme, and Labeling Solution, is incubated at 37°C for 60 minutes in the dark. Stop Solution then terminates the reaction before analysis.
Suitable inputs are suspension cells and adherent cells, with a minimum of 5 × 10^5 cells per sample. The protocol starts with 1 × 10^6 cells per sample. Adherent cultures can release apoptotic cells into the supernatant, so the supernatant should be collected together with the attached cells. Permeabilization must not exceed 15 minutes, because excessive time or temperature breaks cells. Labeling Solution and TdT Enzyme should not be vortexed, and chelating agents such as EDTA should be avoided because they render TdT ineffective.
| Product name: | One-step TUNEL Flow Cytometry Apoptosis Kit (Red, AF647) |
| SKU: | AKES113 |
| Detection method: | Fluorometric method, flow cytometry |
| Sample type: | Cell |
| Sample types supported: | Suspension cells and adherent cells |
| Assay time: | 2.5 to 3 hours |
| Detection instrument: | Flow cytometer |
| Dye type: | AF647 |
| Ex/Em: | 650/670 nm |
| Flow channel: | APC |
| Research use: | Research Use Only |
- TdT-mediated labeling of exposed 3'-OH DNA ends in apoptotic cells
- AF647 dye read in the APC channel (Ex/Em 650/670 nm)
- Suitable for suspension and adherent cells
- Fixation, permeabilization, labeling, and detection in a single workflow
- Approximate assay time of 2.5 to 3 hours
- Research use only
When cells undergo apoptosis, specific DNA endonucleases are activated and cut genomic DNA between nucleosomes. After fixation and permeabilization, the exposed 3'-OH ends of the broken DNA are catalyzed by Terminal Deoxynucleotidyl Transferase (TdT) with fluorescein-labeled dUTP (AF647 in this kit). The incorporated label is detected by flow cytometry in the APC channel.
- Quantification of apoptotic cells in suspension cell cultures by flow cytometry
- Detection of apoptosis in adherent cell cultures, including detached cells collected from the supernatant
- Assessment of DNA fragmentation through TdT labeling of exposed 3'-OH ends
- Evaluation of apoptosis induction conditions, such as drug dose and time course, with appropriate negative controls
| Component | Storage |
| TdT Equilibration Buffer | -20°C |
| TdT Enzyme | -20°C |
| Labeling Solution (AF647) | -20°C, protect from light |
| Fixation Buffer | -20°C |
| Permeabilization Buffer | -20°C |
| Stop Solution | -20°C |
- PBS buffer with 1% BSA, pH 7.2 to 7.4 (user-supplied)
- Flow cytometer with APC channel
- Centrifuge
- Ice bath
- Incubator or heat block set to 37°C
- Room-temperature bench space and 4°C storage
Cells (suspension or adherent)
- Collect the cells and count them. Take 1 × 10^6 cells, centrifuge at 300 × g for 5 min, and discard the supernatant. For adherent cells, collect the supernatant as well, because some apoptotic cells are suspended in it.
- Resuspend the cells in 200 µL of PBS (with 1% BSA). Add 200 µL of Fixation Buffer and incubate at room temperature for 60 min, mixing every 15 min.
- Add 1 mL of PBS (with 1% BSA), centrifuge at 600 × g for 5 min, and discard the supernatant.
- Add 200 µL of Permeabilization Buffer and incubate on ice for 10 min. Thaw the Permeabilization Buffer and place it at 4°C before use. The maximum permeabilization time should not exceed 15 min.
- Add 1 to 2 mL of PBS (with 1% BSA) to stop the reaction, mix by gently pipetting, centrifuge at 600 × g for 5 min, and discard the supernatant.
- Add 200 µL of TdT Equilibration Buffer, mix by gently pipetting, and incubate at 37°C for 15 min. Centrifuge at 600 × g for 5 min, discard the supernatant, and collect the cell pellet.
- Prepare the Labeling Working Solution fresh before use, at 100 µL per sample: 70 µL TdT Equilibration Buffer, 20 µL Labeling Solution, and 10 µL TdT Enzyme for the experimental group. For the negative control, use 80 µL TdT Equilibration Buffer, 20 µL Labeling Solution, and no TdT Enzyme. Gently pipette the Labeling Working Solution to incorporate the TdT Enzyme; do not vortex.
- Incubate at 37°C for 60 min in the dark, mixing gently every 20 min.
- Add 1 mL of Stop Solution, mix gently by pipetting, and incubate at room temperature for 5 min. Add 300 to 500 µL of PBS (with 1% BSA), centrifuge at 600 × g for 5 min, and discard the supernatant.
- Add 200 µL of PBS (with 1% BSA) to resuspend the cells by gently pipetting 10 to 20 times, then analyze by flow cytometry. Detect as soon as possible to avoid fluorescence quenching.
Cell collection and fixation
- Collect cells and count them. Take 1×106 cells and centrifuge at 300×g for 5 min. Discard the supernatant.
- For adherent cultures, collect the cells detached into the supernatant along with the adherent fraction.
- Resuspend cells in 200 µL PBS (with 1% BSA), add 200 µL Fixation Buffer, and incubate at room temperature for 60 min. Mix every 15 min.
- Add 1 mL PBS (with 1% BSA), centrifuge at 600×g for 5 min, and discard the supernatant.
Permeabilization and TdT labeling
- Add 200 µL Permeabilization Buffer and incubate on ice for 10 min. Thaw the buffer and place it at 4°C before use. Do not exceed 15 min of permeabilization.
- Add 1 to 2 mL PBS (with 1% BSA) to stop the reaction, mix gently, centrifuge at 600×g for 5 min, and discard the supernatant.
- Add 200 µL TdT Equilibration Buffer, mix gently, and incubate at 37°C for 15 min.
- Centrifuge at 600×g for 5 min, discard the supernatant, and collect the cell pellet.
- Prepare Labeling Working Solution fresh before use, using 100 µL per sample. Gently pipette the TdT Enzyme into the Labeling Working Solution. Do not vortex.
- Resuspend the cell pellet in the Labeling Working Solution and incubate at 37°C for 60 min in the dark, mixing gently every 20 min.
Stop, wash, and detection
- Add 1 mL Stop Solution, mix gently, and incubate at room temperature for 5 min.
- Add 300 to 500 µL PBS (with 1% BSA), centrifuge at 600×g for 5 min, and discard the supernatant.
- Resuspend cells in 200 µL PBS (with 1% BSA) and analyze by flow cytometry using the APC channel.
- Detect samples as soon as possible to limit fluorescence quenching.
What sample types and cell numbers does the kit support?
The kit detects apoptosis in suspension cells and adherent cells. The minimum number of cells per sample is 5 × 10^5, and the protocol starts with 1 × 10^6 cells.
How should the kit and its components be stored?
Store the kit at -20°C, protected from light, for up to 12 months. Keep the TdT Enzyme strictly at -20°C and return it to storage immediately after use. Store the Labeling Solution in the dark and aliquot it to avoid freeze-thaw cycles.
What should I do if I do not detect apoptotic cells?
Set up gradient concentrations and induction times to find suitable apoptosis induction conditions. Also check cell number and permeabilization, because too many cells or insufficient permeabilization can reduce detection.
Which factors can interfere with the assay?
Chelating agents such as EDTA make TdT ineffective, so use a buffer without chelating agents for washing. Vortexing the Labeling Solution or TdT Enzyme, and permeabilization that is too warm or too long, can damage cells or reduce labeling. Include a negative control without TdT Enzyme in each experiment.
What instrument settings should I use?
Use a flow cytometer and select the APC channel, which matches the AF647 dye with excitation and emission at 650 and 670 nm. Centrifuge is also required for the washing steps.
How long does the assay take?
The assay time is 2.5 to 3 hours, including a 60-minute fixation at room temperature and a 60-minute TdT labeling reaction at 37°C.
| Storage temperature: | -20°C |
| Shelf life: | 12 months |
| Light sensitivity: | Store with shading light protection |
| Shipping: | Ice bag |
Cautions
- For research use only.
- The minimum number of cells for this kit is 5×105.
- Include a negative control without TdT Enzyme in each experiment to confirm TUNEL specificity.
- Do not vortex the Labeling Solution or TdT Enzyme. Avoid freeze-thaw cycles; aliquot reagents into smaller volumes.
- Keep permeabilization brief and cold. Excessive temperature or time (over 15 min) can break cells.
- Use buffers without chelating agents such as EDTA, which inactivate TdT.
- When resuspending cells, pipette gently 10 to 20 times to avoid cell damage and bubbles.
- Carefully remove supernatants after centrifugation to avoid cell loss.
- Insufficient fixation can result in cell debris; fix cells fully before permeabilization.