Description
Phosphate Colorimetric Assay Kit (Malachite Green)
The Phosphate Colorimetric Assay Kit (Malachite Green) quantifies inorganic phosphate in serum, plasma, tissue homogenate, and cell lysate samples. The assay uses a malachite green chromogenic reaction with phosphate and molybdate under acidic conditions, read at 636 nm on a microplate reader. Researchers use it for phosphate quantification in Research Use Only studies of phosphate homeostasis and inorganic ion levels.
Phosphorus is an essential mineral and one of the essential nutrients for all animals. It is required for maintaining phosphate homeostasis, so the phosphate content of biological fluids and tissues reflects a central physiological balance. This kit quantifies phosphate, the analyte that reports phosphorus status in a sample. Results are expressed as phosphate concentration for serum and plasma, or as content per kilogram of wet tissue weight or per 10^6 cells for tissue and cell samples.
Researchers use this kit in the listed research areas of Inorganic Ions and Oxidative Stress to compare phosphate content across sample types. The supporting documentation reports testing of rat liver, heart, spleen, brain, and kidney homogenates, mouse heart, lung, and ovarian homogenates, rat serum, rat plasma, human plasma, and HL-40 cells. The kit is for research use only and is not intended for clinical diagnosis.
The malachite green method relies on phosphoric acid and molybdate reacting under acidic conditions to form compounds that produce colored substances with malachite green. Color depth is proportional to phosphate content, and phosphate is determined indirectly by measuring optical density at 636 nm on a microplate reader set to 630-650 nm. Standards or samples (20 µL) are combined with 200 µL of freshly prepared chromogenic working solution. The mixture is shaken for 5 s and incubated at 37 °C for 20 min with shading from light before reading.
Serum and plasma are deproteinized with an equal volume of protein precipitator, then centrifuged to obtain a supernatant for testing. Tissue samples are washed in cold PBS, homogenized in normal saline with a dounce homogenizer at 4 °C, and centrifuged at 12000×g to remove insoluble material. Cell samples are washed in PBS, disrupted by ultrasonication in normal saline at 4 °C, and clarified by the same centrifugation step. Supernatants are kept on ice until detection and are prepared for testing with the same protein precipitator step.
Reported performance includes an analytical sensitivity of 0.001 mmol/L and a detection range of 0.001-0.20 mmol/L. Intra-assay precision was 3.8-4.2% CV across three human serum samples, and inter-assay precision was 7.8-8.3% CV across the same samples assayed by three operators. Mean recovery across low, middle, and high concentration samples was 98%. The product page lists intra-assay CV as 4% and inter-assay CV as 8%.
| Product Name: | Phosphate Colorimetric Assay Kit (Malachite Green) |
| SKU: | MAES0245 |
| Assay Type: | Quantitative, colorimetric |
| Sample Type: | Serum, plasma, tissue, and cell samples |
| Sample Volume: | 20 µL |
| Assay Time: | 35 min |
| Detection Instrument: | Microplate reader (630-650 nm; optimum 636 nm) |
| Detection Range: | 0.001-0.20 mmol/L |
| Research Area: | Inorganic ions; oxidative stress |
| Sensitivity: | 0.001 mmol/L |
| Detection Range: | 0.001-0.20 mmol/L |
| Intra-assay CV: | 4% (3.8-4.2% across three human serum samples, n=20 replicates each) |
| Inter-assay CV: | 8% (7.8-8.3% across three human serum samples, 20 runs in duplicate, three operators) |
| Recovery: | 98% average (96-99% across three standard levels) |
- Quantitative malachite green colorimetric method for phosphate
- Detection range 0.001-0.20 mmol/L; analytical sensitivity 0.001 mmol/L
- Sample types: serum, plasma, tissue homogenate, and cell samples
- Total assay time 35 min; 20 µL sample per well
- Absorbance read at 636 nm (instrument range 630-650 nm)
Phosphoric acid and molybdate form a complex under acidic conditions, which forms colored species with malachite green. The color intensity is proportional to phosphate content. Phosphate is calculated indirectly from the absorbance at 636 nm, after subtracting the blank and interpolating from a standard curve.
- Quantifying phosphate in serum and plasma samples in phosphate homeostasis studies
- Measuring phosphate in tissue homogenates from rodent organs
- Measuring phosphate in cell lysates for inorganic ion research
- Assessing phosphate levels in oxidative stress research workflows
| Component | Storage |
| Chromogenic Agent A | 2-8°C, 12 months, protect from light |
| Chromogenic Agent B | 2-8°C, 12 months |
| Protein Precipitator | 2-8°C, 12 months |
| Standard Solution (10 mmol/L) | 2-8°C, 12 months |
| Microplate | No requirement |
| Plate Sealer | No requirement |
| Sample Layout Sheet | No requirement |
- Microplate reader capable of 630-650 nm reading (636 nm optimum)
- Micropipettor and sterile tips
- Incubator set to 37°C
- Vortex mixer
- Centrifuge capable of 12000×g, including 4°C operation
- Normal saline (0.9% NaCl), used for sample homogenization and dilution
- PBS (0.01 M, pH 7.4), used for tissue and cell washing
- Double-distilled water, used for working solution and standard dilutions
- Dounce homogenizer (tissue) or ultrasonic cell disruptor (cells)
- 96-well microplate and plate sealer
Serum/plasma
- For each well, add 10 µL of sample and 10 µL of protein precipitator, then mix well.
- Centrifuge at 12000×g for 10 minutes, collect the supernatant, and keep it on ice for detection.
- Dilute samples before use with normal saline (0.9% NaCl). Recommended dilution factors: rat serum 8-12, rat plasma 8-12, human plasma 8-12.
- Dilution factors are for reference only. For other sample types, perform a pretest to confirm the dilution factor.
Tissue
- Harvest the amount of tissue needed for each assay (initial recommendation 20 mg).
- Wash tissue in cold PBS (0.01 M, pH 7.4).
- Homogenize 20 mg tissue in normal saline (0.9% NaCl) with a dounce homogenizer at 4 °C.
- Centrifuge at 12000×g for 10 minutes to remove insoluble material. Collect the supernatant and keep it on ice.
- For each well, add 10 µL of supernatant and 10 µL of protein precipitator, then mix well.
- Centrifuge at 12000×g for 10 minutes, collect the supernatant, and keep it on ice for detection.
- Recommended dilution factors for 10% homogenates: rat liver, heart, spleen, brain, and kidney 15-25; mouse heart and lung 15-25; mouse ovarian 50-70.
- Dilution factors are for reference only. For other tissue types, perform a pretest to confirm the dilution factor.
Cells (adherent or suspension)
- Harvest the number of cells needed for each assay (initial recommendation 1×10^6 cells).
- Wash cells with PBS (0.01 M, pH 7.4).
- Homogenize 1×10^6 cells in 200 µL normal saline (0.9% NaCl) with an ultrasonic cell disruptor at 4 °C.
- Centrifuge at 12000×g for 10 minutes at 4 °C to remove insoluble material. Collect the supernatant and keep it on ice.
- For each well, add 10 µL of supernatant and 10 µL of protein precipitator, then mix well.
- Centrifuge at 12000×g for 10 minutes, collect the supernatant, and keep it on ice for detection.
- Recommended dilution factor for 10^6 HL-40 cells: 2-8. Use normal saline (0.9% NaCl) as the diluent.
- Dilution factors are for reference only. For other cell types, perform a pretest to confirm the dilution factor.
Reagent preparation
- Equilibrate all kit reagents to room temperature before use.
- Prepare chromogenic working solution fresh for each well: mix 75 µL Chromogenic Agent A, 25 µL Chromogenic Agent B, and 100 µL double-distilled water (200 µL per well). Incubate at 37°C for 1 h. Use the same day.
- Prepare 0.2 mmol/L standard by diluting 20 µL of 10 mmol/L Standard Solution with 980 µL double-distilled water. Prepare fresh; store at 2-8°C for up to 7 days if needed.
- Prepare the standard curve fresh for each assay by diluting the 0.2 mmol/L standard with double-distilled water to 0.20, 0.15, 0.12, 0.10, 0.05, 0.02, 0.01, and 0 mmol/L (0 µL to 200 µL of standard, with 200 µL to 0 µL water).
Sample preparation
- Serum or plasma: per well, add 10 µL sample and 10 µL Protein Precipitator, mix well. Centrifuge at 12000×g for 10 min. Collect supernatant for the assay.
- Tissue: homogenize 20 mg tissue in normal saline (0.9% NaCl) with a Dounce homogenizer at 4°C after washing in cold PBS. Centrifuge at 12000×g for 10 min and collect supernatant. Per well, add 10 µL supernatant and 10 µL Protein Precipitator, mix, centrifuge at 12000×g for 10 min, and collect supernatant.
- Cells (adherent or suspension): harvest 1×10^6 cells, wash with PBS, and homogenize in 200 µL normal saline with an ultrasonic cell disruptor at 4°C. Centrifuge at 12000×g for 10 min at 4°C and collect supernatant. Per well, add 10 µL supernatant and 10 µL Protein Precipitator, mix, centrifuge at 12000×g for 10 min, and collect supernatant.
- Dilute samples as needed with normal saline (0.9% NaCl). Recommended dilution factors (reference only): 10% rat liver, heart, spleen, brain, or kidney homogenate and 10% mouse heart or lung homogenate, 15-25; 10% mouse ovarian homogenate, 50-70; rat serum, rat plasma, and human plasma, 8-12; 10^6 HL-40 cells, 2-8. For other sample types, run a pretest to confirm the dilution factor.
Assay
- Add 20 µL of each standard or sample to the wells of the microplate.
- Add 200 µL chromogenic working solution to each well.
- Mix thoroughly on the microplate reader for 5 s, then incubate at 37°C for 20 min in the dark.
- Read the OD of each well at 636 nm.
Calculation
- Average duplicate readings for each standard. Subtract the mean blank (standard #1, 0 mmol/L) OD from all readings to obtain the absolute OD.
- Plot absolute OD (y) against standard concentration in mmol/L (x) and fit a linear standard curve (y = ax + b).
- Serum or plasma: Phosphate (mmol/L) = (ΔA − b) ÷ a × f × 2, where ΔA = OD sample − OD blank, f = pre-assay dilution factor, and 2 = dilution factor during supernatant preparation.
- Tissue: Phosphate (mmol/kg wet weight) = (ΔA − b) ÷ a ÷ (m ÷ V) × f × 2, where m = tissue weight (g) and V = volume of normal saline used in preparation (mL).
- Cells: Phosphate (mmol/10^6 cells) = (ΔA − b) ÷ a ÷ (n ÷ V) × f × 2, where n = number of cells (×10^6).
Standard curve: y = ax + b, where y is the absolute OD value and x is standard concentration (mmol/L). Serum/plasma: Phosphate (mmol/L) = (ΔA − b) ÷ a × f × 2*. Tissue: Phosphate (mmol/kg wet weight) = (ΔA − b) ÷ a ÷ (m ÷ V) × f × 2*. Cells: Phosphate (mmol/10^6) = (ΔA − b) ÷ a ÷ (n ÷ V) × f × 2*.
- ΔA = OD(Sample) − OD(Blank). Absolute OD for standards is the standard OD minus the mean OD of the blank (Standard #1).
- a and b are the slope and intercept of the standard curve, y = ax + b, generated from absolute OD values and standard concentrations, using graphing software or Excel.
- m is the weight of tissue sample in g.
- V is the volume of normal saline used in the sample preparation step, in mL.
- n is the number of cells in the sample, in units of 10^6 cells.
- f is the dilution factor of the sample before testing.
- 2* is the dilution factor of the sample during supernatant preparation, which is 2 because 10 µL of sample is combined with 10 µL of protein precipitator.
- Worked example from the manual: for 10% rat liver homogenate diluted 20-fold, with standard curve y = 5.1343x − 0.0073, blank OD 0.142, and sample OD 0.564, phosphate was 30.06 mmol/kg wet weight, calculated with m = 0.1 g, V = 0.9 mL, f = 20, and 2* = 2.
What sample types can be tested with this kit?
The kit measures phosphate in serum, plasma, tissue, and cell samples. Serum and plasma are processed with protein precipitator, while tissue and cell samples are homogenized in normal saline before the same precipitation step.
How should reagents be stored, and how long are they stable?
Kit components should be stored at 2-8 °C for 12 months, with the chromogenic agent A protected from light. Working chromogenic solution must be prepared on the spot and used within the same day, and 0.2 mmol/L standard can be stored at 2-8 °C for 7 days. Reagents from different kits must not be mixed.
What if my sample is outside the detection range or the OD is too high?
The detection range is 0.001-0.20 mmol/L. If the absolute OD value is more than 1.2, increase the dilution ratio and test again. If the concentration is not within the detection range, perform an extra dilution or concentration on the sample.
What key points affect assay performance?
Incubate chromogenic agent A at 37 °C until no gelatinous substance remains before use. Avoid phosphorus contamination by washing experimental equipment several times (about 10 times). Equilibrate all reagents to room temperature before use.
What instrument settings are required?
Use a microplate reader set to 630-650 nm, with the optimum wavelength of 636 nm for readout. After adding 200 µL of chromogenic working solution to each well, mix for 5 s and incubate at 37 °C for 20 min with shading from light before reading.
How long does the assay take?
The product page lists an assay time of 35 minutes. Chromogenic working solution must first be incubated at 37 °C for 1 hour, and the plate is then incubated at 37 °C for 20 minutes with shading before reading at 636 nm.
| Storage Temperature: | 2-8°C |
| Light Protection: | Store with shading light protection |
| Shelf Life: | 12 months |
| Working Solution: | Prepare fresh on the day of use; use within the same day |
| Standard Solution (10 mmol/L): | 2-8°C, 12 months |
Notes
- For Research Use Only. Not for use in clinical diagnosis or therapeutic procedures.
- Avoid phosphate contamination. Wash labware several times (about 10 times) before use.
- If the absolute OD exceeds 1.2, increase the dilution factor and repeat the assay.
- Incubate Chromogenic Agent A at 37°C until no gelatinous substance remains before use.
- Prepare chromogenic working solution and standard dilutions fresh; discard unused working standards.
- Centrifuge small reagent volumes before use to recover sufficient reagent.
- Do not mix reagents from different kit lots.
- Wear a lab coat and gloves. Include a pretest for sample types not listed in the protocol.