Description
Triglyceride Fluorometric Assay Kit
The Assay Genie Triglyceride Fluorometric Assay Kit (MAES0257) quantifies triglyceride (TG) in serum, plasma, tissue and cell samples. TG is converted by enzymes to hydrolytic products, which are then catalyzed to fluorescent substances read at Ex/Em 535/587 nm. Researchers use the kit for lipid and energy metabolism studies where TG must be measured in small sample volumes. The kit is for Research Use Only.
Triglyceride (TG) is the analyte measured by this kit, and the product is listed under the glycolysis and lipid metabolism and energy metabolism research areas. The kit quantifies TG content in serum, plasma, tissue, and cell samples. Results are reported in mmol/L for serum and plasma and in µmol/g protein for tissue and cell lysates, so the unit depends on the sample type. The assay is also known by the synonym TG. The kit is labeled for research use only and is not intended for clinical diagnosis.
Researchers measure triglyceride to study lipid and energy metabolism. The sources describe use across several physiology models: porcine heart and liver tissue homogenates, CHO and 293T cells, and human and mouse serum. Comparing TG content across these matrices allows investigators to examine lipid handling in different tissues, cell lines, and species. The manual states that the kit is intended for determining triglyceride in serum, plasma, tissue, and cell samples, and it must be used for research purposes only.
The assay is enzymatic and fluorometric. Triglyceride is converted by enzymes into hydrolytic products, which are then catalyzed by enzymes to produce fluorescent substances. For each well, a 200 µL chromogenic working solution is prepared on the day of use by mixing 196 µL of Enzyme Working Solution with 4 µL of Probe, and it must be protected from light. Then 20 µL of standard or sample is added to the appropriate well, followed by 200 µL of working solution. After mixing, the plate is incubated at 37°C for 5 min. Fluorescence is read at an excitation wavelength of 535 nm and an emission wavelength of 587 nm.
Serum and plasma can be assayed directly without extraction. Tissue and cell samples are homogenized in extraction solution, clarified by centrifugation at 10000 × g for 10 min, and the supernatant is kept on ice until assayed. Tissue homogenization uses 20 mg of tissue in 180 µL extraction solution at 4°C. Cell homogenization uses 1 × 10^6 cells in 200 µL extraction solution with an ultrasonic cell disruptor at 4°C. The protein concentration of each supernatant must be determined separately so that tissue and cell results can be normalized.
Reported performance includes an analytical sensitivity of 0.8 µmol/L and a detection range of 0.8 to 30 µmol/L. In the manual's replicate study, intra-assay precision ranged from 3.7% to 4.2% CV and inter-assay precision ranged from 8.4% to 10.0% CV. Mean recovery was 105%, with individual recoveries of 103% to 107% across three spiked concentrations. The kit specifications give an intra-assay CV of 3.9% and an inter-assay CV of 9.1%.
| Product Name: | Triglyceride Fluorometric Assay Kit |
| SKU: | MAES0257 |
| Synonyms: | TG |
| Sample Type: | Serum, plasma, tissue and cell samples |
| Assay Type: | Quantitative |
| Detection Method: | Fluorometric |
| Detection Instrument: | Fluorescence microplate reader (Ex/Em 535 nm/587 nm) |
| Sensitivity: | 0.8 µmol/L |
| Detection Range: | 0.8-30 µmol/L |
| Intra-assay CV: | 3.9% |
| Inter-assay CV: | 9.1% |
| Valid Period: | 12 months |
| Detection Method: | Fluorometric |
| Excitation/Emission: | 535 nm/587 nm |
| Sample Volume: | 20 µL |
| Sensitivity: | 0.8 µmol/L |
| Detection Range: | 0.8-30 µmol/L |
| Intra-assay CV (three human serum samples, n=20): | 3.7-4.2% |
| Inter-assay CV (three human serum samples, 20 runs in duplicate): | 8.4-10.0% |
| Mean recovery: | 105% (range 103-107%) |
- Quantitative fluorometric detection of triglyceride (Ex/Em 535 nm/587 nm)
- Detection range 0.8-30 µmol/L with sensitivity of 0.8 µmol/L
- Sample types: serum, plasma, tissue and cell samples
- Requires 20 µL of sample per well
- Intra-assay CV 3.9% and inter-assay CV 9.1%
- Standard solution of 1 mmol/L included for standard curve preparation
Triglyceride is converted by enzyme to produce hydrolytic products, which is catalyzed by enzymes to produce fluorescent substances. The content of triglyceride in samples can be calculated by measuring the fluorescence value.
- Measurement of triglyceride levels in serum and plasma samples for lipid metabolism research
- Quantification of triglyceride in tissue homogenates, normalized to protein concentration
- Measurement of triglyceride content in cell lysates from adherent or suspension cultures
- Energy metabolism studies requiring fluorometric TG quantification
| Component | Storage |
| Enzyme Working Reagent 1 | 2-8°C, protect from light |
| Extraction Reagent 2 (Extraction Solution) | 2-8°C |
| Reagent 3 (Probe) | -20°C, protect from light |
| Reagent 4 (1 mmol/L Standard Solution) | 2-8°C |
| Black microplate | No requirement |
- Fluorescence microplate reader capable of Ex/Em 535 nm/587 nm
- Centrifuge capable of 10000 × g
- Incubator set to 37°C
- Ice bath
- PBS (0.01 M, pH 7.4) for tissue and cell washing
- Dounce homogenizer (tissue) or ultrasonic cell disruptor (cells)
- Protein concentration assay for normalizing tissue and cell lysates
- Precision pipettes and tips, microcentrifuge tubes, and a 96-well plate
Serum/plasma
- Detect serum or plasma directly.
- If samples are not detected on the same day, store serum or plasma at -80°C for up to one month.
- Sample volume is 20 µL per well.
- For human serum and mouse serum, the recommended dilution factor is 100-120.
- Dilute samples with extraction solution. For other sample types, perform a pretest to confirm the dilution factor.
Tissue
- Harvest the amount of tissue needed for each assay (initial recommendation: 20 mg).
- Wash the tissue in cold PBS (0.01 M, pH 7.4).
- Homogenize 20 mg of tissue in 180 µL of extraction solution with a homogenizer at 4°C.
- Centrifuge at 10000 × g for 10 min to remove insoluble material. Collect the supernatant and keep it on ice for detection.
- Determine the protein concentration of the supernatant using a separate protein assay.
- Dilution factors (use extraction solution as diluent): 10% porcine heart tissue homogenate, 3-6; 10% porcine liver tissue homogenate, 8-10. For other tissue types, perform a pretest to confirm the dilution factor.
Cells (adherent or suspension)
- Harvest the number of cells needed for each assay (initial recommendation: 1 × 10^6 cells).
- Wash cells with PBS (0.01 M, pH 7.4).
- Homogenize 1 × 10^6 cells in 200 µL of extraction solution with an ultrasonic cell disruptor at 4°C.
- Centrifuge at 10000 × g for 10 min to remove insoluble material. Collect the supernatant and keep it on ice for detection.
- Determine the protein concentration of the supernatant using a separate protein assay.
- Dilution factors (use extraction solution as diluent): 1 × 10^6 CHO cells, 1-2; 1 × 10^6 293T cells, 1-2. For other cell types, perform a pretest to confirm the dilution factor.
Reagent preparation
- Equilibrate all reagents to room temperature before use.
- Prepare chromogenic working solution fresh for each use: per well, mix 196 µL Enzyme Working Reagent 1 with 4 µL Probe (Reagent 3) to a total of 200 µL. Protect from light.
- Prepare 30 µmol/L standard solution by diluting 30 µL of 1 mmol/L Standard Solution with 970 µL Extraction Reagent 2. Store at 2-8°C for up to 3 days.
- Prepare a fresh standard curve for each assay by diluting the 30 µmol/L standard with Extraction Reagent 2 as follows: 0, 6, 12, 18, 24 and 30 µmol/L using 0, 40, 80, 120, 160 and 200 µL of 30 µmol/L standard with 200, 160, 120, 80, 40 and 0 µL of Extraction Reagent 2, respectively.
Sample preparation
- Serum and plasma: test directly. If not tested the same day, store at -80°C for up to one month.
- Tissue: wash tissue in cold PBS. Homogenize 20 mg tissue in 180 µL Extraction Reagent 2 at 4°C using a Dounce homogenizer. Centrifuge at 10000 × g for 10 min. Keep the supernatant on ice for testing.
- Cells (adherent or suspension): wash 1 × 10^6 cells with PBS. Homogenize in 200 µL Extraction Reagent 2 at 4°C using an ultrasonic cell disruptor. Centrifuge at 10000 × g for 10 min. Keep the supernatant on ice for testing.
- Determine protein concentration of tissue and cell supernatants for normalization.
- Dilute samples as needed. Recommended dilution factors (reference only): 10% porcine heart tissue homogenate 3-6; 10% porcine liver tissue homogenate 8-10; 1 × 10^6 CHO cells 1-2; 1 × 10^6 293T cells 1-2; human serum 100-120; mouse serum 100-120. Use Extraction Reagent 2 as diluent. Pretest other sample types to confirm dilution.
Assay procedure
- Add 20 µL of each standard to standard wells and 20 µL of each sample to sample wells.
- Add 200 µL of chromogenic working solution to each well.
- Mix thoroughly and incubate at 37°C for 5 min.
- Measure fluorescence at Ex 535 nm and Em 587 nm.
Calculation
- Average the duplicate readings for each standard.
- Subtract the mean blank (standard #1, 0 µmol/L) reading from all standard readings to obtain absolute fluorescence values.
- Plot absolute fluorescence (y) against concentration (x) and fit a linear standard curve (y = ax + b).
- Serum/plasma: TG (mmol/L) = (ΔF − b) ÷ a × f ÷ 1000, where ΔF is sample fluorescence minus blank, and f is the dilution factor.
- Tissue/cell: TG (µmol/g protein) = (ΔF − b) ÷ a ÷ Cpr × f, where Cpr is protein concentration in g protein/L.
Serum (plasma): TG content (mmol/L) = (ΔF - b) / a × f ÷ 1000; Tissue and cells: TG content (µmol/g prot) = (ΔF - b) / a ÷ Cpr × f
- ΔF is the absolute fluorescence intensity of the sample (F sample - F blank).
- a and b are the slope and intercept of the standard curve, y = ax + b, where y is the absolute fluorescence value of the standard and x is its concentration in µmol/L.
- Cpr is the protein concentration of the sample in gprot/L, determined separately from the supernatant.
- f is the dilution factor of the sample before testing. Multiply by f after dividing by the standard curve values.
- 1000 converts µmol to mmol for serum and plasma results.
- Standard curve: average the duplicate readings for each standard, then subtract the mean blank (Standard 1, 0 µmol/L) from all standard readings. Standards are 0, 6, 12, 18, 24, and 30 µmol/L.
- Worked example from the manual: for human serum diluted 100-fold, standard curve y = 152.95x - 74.284, sample average fluorescence 2664, and control average fluorescence 1316, TG content = (2664 - 1316 + 74.284) ÷ 152.95 × 100 ÷ 1000 = 0.93 mmol/L.
Which sample types can be tested with this kit?
The kit is intended for triglyceride determination in serum, plasma, tissue, and cell samples, including adherent and suspension cells. Each well requires 20 µL of sample.
How should samples and kit components be stored?
Serum and plasma should be assayed the same day or stored at -80°C for up to one month. Kit components are stored for 12 months, with the Probe at -20°C and the other components at 2-8°C, all protected from light.
What should I do if my results fall outside the detection range?
The detection range is 0.8 to 30 µmol/L. If a sample is not within this range, perform an extra dilution or concentration of the sample before testing.
What key points affect assay performance?
Clean the preparation equipment several times before preparing the chromogenic working solution to prevent contamination by impurities. Prepare the working solution on the day of use and protect it from light. Do not mix components from different kits.
What instrument settings are required?
Use a fluorescence microplate reader with excitation at 535 nm and emission at 587 nm. Read the plate after incubation at 37°C for 5 min, and use a black microplate.
How long does the assay take?
The assay time is 30 min, which includes the 5 min incubation at 37°C after adding the working solution.
| Enzyme Working Reagent 1: | 2-8°C, protect from light |
| Extraction Reagent 2: | 2-8°C |
| Reagent 3 (Probe): | -20°C, protect from light |
| Reagent 4 (1 mmol/L Standard Solution): | 2-8°C |
| Valid period: | 12 months |
| Serum/plasma sample storage before testing: | -80°C for up to one month |
Notes
- For Research Use Only. Not for use in clinical diagnosis or therapeutic procedures.
- Protect the Probe, Enzyme Working Reagent and chromogenic working solution from light.
- Prepare the chromogenic working solution on the day of use.
- Do not mix reagents from different kits.
- Centrifuge small reagent volumes before use to ensure sufficient volume is recovered.
- Clean preparation equipment thoroughly before preparing the chromogenic working solution to avoid impurity contamination.
- If a sample is outside the detection range, dilute or concentrate the sample and repeat the assay.
- Pretest samples not listed in the manual to confirm dilution factor and performance.
- Wear a lab coat and gloves during use.