Description
system_update_altDatasheet
| Product Name: | GeniePlex Sepsis Extended 14-Plex (96 Tests) |
| Product Code: | HUAMPM058 |
| Price: | €3,629 (96 Tests) |
| Size: | 96 Tests |
| Reactivity: | Human |
| Analytes (14): | TNF-alpha, IL-18/IGIF, IL-12p70, CXCL10/IP-10, IL-1beta, IL-2, IL-1RA/IL-1RN, IL-8/CXCL8, IFN-gamma, TGF-beta1, CCL2/MCP-1, IL-10/CSIF, CCL3/MIP-1-alpha, and IL-6 |
| Sensitivity (LODs): | TNF-alpha: N/A pg/mL, IL-18/IGIF: N/A pg/mL, IL-12p70: < 3 pg/mL, CXCL10/IP-10: < 3 pg/mL, IL-1beta: < 5 pg/mL, IL-2: < 3 pg/mL, IL-1RA/IL-1RN: < 20 pg/mL, IL-8/CXCL8: < 5 pg/mL, IFN-gamma: < 3 pg/mL, TGF-beta1: < 20 mUnit/mL pg/mL, CCL2/MCP-1: < 2 pg/mL, IL-10/CSIF: < 2 pg/mL, CCL3/MIP-1-alpha: < 2 pg/mL, IL-6: < 5 pg/mL |
| Sample types: | Cell culture supernatant, serum, plasma, bodily fluid and tissue/cell lysate |
| Assay Type: | Multiplex |
| Detection Method: | Flow Cytometry |
| Assay Time: | 2 hours |
| 1x Ab-conjugated beads (premixed): | anti-human TNF-alpha, anti-human IL-18/IGIF, S4P12 - anti-human IL-12p70, S4P4 - anti-human CXCL10/IP-10, S4P6 - anti-human IL-1beta, S4P7 - anti-human IL-2, S4P8 - anti-human IL-1RA/IL-1RN, S4P9 - anti-human IL-8/CXCL8, S4P9 - anti-human IFN-gamma, S4P9 - anti-human TGF-beta1, S5P1 - anti-human CCL2/MCP-1, S5P10 - anti-human IL-10/CSIF, S5P2 - anti-human CCL3/MIP-1-alpha, S5P6 - anti-human IL-6 |
| 2x Biotin-detection Ab (premixed): | Biotin - anti-human TNF-alpha, Biotin - anti-human IL-18/IGIF, Biotin - anti-human IL-12p70, Biotin - anti-human CXCL10/IP-10, Biotin - anti-human IL-1beta, Biotin - anti-human IL-2, Biotin - anti-human IL-1RA/IL-1RN, Biotin - anti-human IL-8/CXCL8, Biotin - anti-human IFN-gamma, Biotin - anti-human TGF-beta1, Biotin - anti-human CCL2/MCP-1, Biotin - anti-human IL-10/CSIF, Biotin - anti-human CCL3/MIP-1-alpha, Biotin - anti-human IL-6 |
| Buffers: | 2x NR dAb Diluent. One vial containing 1.5 mL of biotin detection antibody diluent. |
| Lyophilized Standard Mix: | Lyophilized Standard Mix- Human Sepsis Extended 14-Plex. Vials A, B and C containing lyophilized recombinant human TNF-alpha, IL-18/IGIF, IL-12p70, CXCL10/IP-10, IL-1beta, IL-2, IL-1RA/IL-1RN, IL-8/CXCL8, IFN-gamma, TGF-beta1, CCL2/MCP-1, IL-10/CSIF, CCL3/MIP-1-alpha, IL-6. |
| LLOQs: | TNF-alpha: N/A pg/mL, IL-18/IGIF: N/A pg/mL, IL-12p70: < 10 pg/mL, CXCL10/IP-10: < 10 pg/mL, IL-1beta: < 10 pg/mL, IL-2: < 10 pg/mL, IL-1RA/IL-1RN: < 40 pg/mL, IL-8/CXCL8: < 10 pg/mL, IFN-gamma: < 10 pg/mL, TGF-beta1: < 50 mUnit/mL pg/mL, CCL2/MCP-1: < 5 pg/mL, IL-10/CSIF: < 5 pg/mL, CCL3/MIP-1-alpha: < 5 pg/mL, IL-6: < 10 pg/mL. |
| ULOQs: | TNF-alpha: N/A pg/mL, IL-18/IGIF: N/A pg/mL, IL-12p70: > 2,000 pg/mL, CXCL10/IP-10: > 2,000 pg/mL, IL-1beta: > 5,000 pg/mL, IL-2: > 2,000 pg/mL, IL-1RA/IL-1RN: > 15,000 pg/mL, IL-8/CXCL8: > 5,000 pg/mL, IFN-gamma: > 2,000 pg/mL, TGF-beta1: > 10,000 mUnit/mL pg/mL, CCL2/MCP-1: > 5,000 pg/mL, IL-10/CSIF: > 5,000 pg/mL, CCL3/MIP-1-alpha: > 2,000 pg/mL, IL-6: > 5,000 pg/mL. |
| Standard dose recovery: | 70-130% |
| Intra-assay CV: | Intra-assay CV: <10% |
| Inter-assay CV: | Inter-assay CV: <20% |
| Cross-reactivity of analytes in the Panel: | Negligible |
| Sample volume: | 15 µL/test |
*Note: The below protocol is a sample protocol. Protocols are specific to each batch/lot. For the correct instructions please follow the protocol included in your kit.
A filter plate washer is required for the following protocol. For more information on GeniePlex click here.
| Step | Protocol |
| 1. | Prepare the filter plate template. |
| 2. | Vortex working bead suspension for 15 seconds. |
| 3. | Add 45 µL of capture bead working suspension to each well. |
| 4. | Remove buffer in the wells by using the Filter Plate Washer connected to a vacuum source. |
| 5. | Remove buffer in the wells by using the Filter Plate Washer connected to a vacuum source. |
| 6. | Add 30 µL of CCS, SPB or TL Assay Buffer to each sample well. |
| 7. | Add 15 µL of samples to each sample well. Add 45 µL of standards to each standard well. Cover the plate with a plate seal. |
| 8. | Incubate on the shaker (set at 700 rpm) for 60 min at room temperature. Protect from light. |
| 9. | Remove the plate seal. Remove solutions in the wells by using the Filter Plate Washer. |
| 10. | Remove solutions in the wells by using the Filter Plate Washer. |
| 11. | Wash the wells three times with 100µL 1x Wash Buffer. |
| 12. | Gently tap the plate bottom onto paper towels to remove residual buffer. |
| 13. | Add 25µL of biotinylated antibody working solution to each well. Cover with plate seal. |
| 14. | Incubate on the shaker (700 rpm) for 30 min at room temperature. Protect from light. |
| 15. | Remove plate seal. Remove solutions. Wash three times with 100 µL 1x Wash Buffer. |
| 16. | Gently tap the plate bottom onto paper towels. |
| 17. | Add 25µL of streptavidin-PE working solution to each well. Cover with plate seal. |
| 18. | Incubate on the shaker (700 rpm) for 20 min at room temperature. Protect from light. |
| 19. | Remove plate seal. Remove solutions. Wash twice with 100 µL 1x Wash Buffer. |
| 20. | Add 150-300µL of 1x Reading Buffer to each well to re-suspend beads. Cover with plate seal. |
| 21. | Shake for 30 seconds at 700 rpm. Read on a flow cytometer. |
GeniePlex Technology
GeniePlex uses a unique mix of antibody-coated encoded microparticles providing an ultra-sensitive technology for the quantitation of analytes.