ELISA Protocol - Step by Step
Workflow Overview
Figure 1. Overview of a Sandwich ELISA protocol workflow
Sample Preparation
General considerations:
According to best practices, extract protein & perform the experiment as soon as possible after sample collection. Alternatively, store the extracts at the designated temperature (-20°C/-80°C). For optimal results, avoid repeated freeze-thaw cycles.
Figure 2. Preparation of serum samples.
Serum:
If using serum separator tubes, allow samples to clot for 30 minutes at room temperature. Centrifuge for 20 minutes at 1,000x g. Collect the serum fraction and assay promptly or aliquot and store the samples at -80°C. Avoid multiple freeze-thaw cycles.
If serum separator tubes are not being used, allow samples to clot overnight at 2-8°C. Centrifuge for 20 minutes at 1,000x g. Remove serum and assay promptly or aliquot and store the samples at -80°C. Avoid multiple freeze-thaw cycles.
Plasma:
Collect plasma using EDTA or heparin as an anticoagulant. Centrifuge samples at 4°C for 15 mins at 1000 × g within 30 mins of collection. Collect the plasma fraction and assay promptly or aliquot and store the samples at -80°C. Avoid multiple freeze-thaw cycles.
Figure 3. Preparation of suspension cell lysates.
Suspension Cell Lysate:
Centrifuge at 2500 rpm for 5 minutes (2-8°C). Add pre-cooling PBS and gently mix. Recollect cells by centrifugation. Add 0.5-1 ml RIPA lysis buffer (NP-40 lysis buffer or Triton X-100 surfactant are not recommended due to their interference with the antigen-antibody reaction). Add a suitable protease inhibitor (e.g. PMSF, working concentration: 1 mmol/L). Lyse cells on ice for 30 min-1 h. During the lysate process, use the pipette tip or intermittent shaking of the centrifuge tube to completely lyse the protein. Alternatively, subject cells to fragmentation by ultrasonic generator (14 μM for 30 s) or ultrasonic cell disruptor (300 W, 3~5 s/time. 30 s intervals, four-five times). At the end of the lysate or ultrasonic disruption, centrifuge at 10000 rpm for 10 minutes (2-8°C). Add the supernatant into an EP tube and store (-80°C).
Figure 4. Preparation of adherent cell lysates.
Adherent Cell Lysate:
Absorb the supernatant and add pre-cooling PBS. Add 0.5-1 ml RIPA lysis buffer (NP-40 lysis buffer or Triton X-100 surfactant are not recommended due to their interference with the antigen-antibody reaction). Add a suitable protease inhibitor (e.g. PMSF, working concentration: 1 mmol/L). Scrape the adherent cells gently with a cell scraper. Add the cell suspension into the centrifuge tube. Lyse cells on ice for 30 min-1 h. During the lysate process, use the pipette tip or intermittent shaking of the centrifuge tube to completely lyse the protein. Alternatively, subject cells to fragmentation by ultrasonic generator (14 μM for 30 s) or ultrasonic cell disruptor (300 W, 3~5 s/time. 30 s intervals, four-five times). At the end of the lysate or ultrasonic disruption, centrifuge at 10000 rpm for 10 minutes (2-8°C). Add the supernatant into an EP tube and store (-80°C).
Figure 5. Preparation of tissue homogenates.
Tissue homogenates:
Rinse tissue with 1X PBS to remove excess blood. Mince tissue after weighing it and homogenize in PBS (the volume depends on the weight of the tissue. 9mL PBS (including protease inhibitors) would be appropriate for 1 gram of tissue. To further lyse cells, you can sonicate the suspension with an ultrasonic cell disrupter or subject it to freeze-thaw cycles. The homogenates are then centrifuged for 5 minutes at 5000 x g and the supernatant is removed for assaying. The total protein concentration was determined by BCA kit and the total protein concentration of each pore sample should not exceed 0.3mg.
Standard Diluton:
1). Add 1ml of Sample dilution buffer into one Standard tube (labelled as Stock Solution), keep the tube at room temperature for 10 min and mix thoroughly.
2). Label 7 Eppendorf tubes with 1/2, 1/4, 1/8, 1/16, 1/32, 1/64 and blank respectively. Aliquot 300μl of the Standard dilution buffer into each tube. Add 300μl of the above (Stock Solution) standard solution into 1st tube and mix thoroughly. Transfer 300μl from 1st tube to 2nd tube and mix thoroughly. Transfer 300μl from 2nd tube to 3rd tube and mix thoroughly, and so on.
Sample dilution buffer is used as blank control.
Figure 6. Standard dilution workflow.
Preparation of biotin-labelled antibody working solution:
Prepare within 1 hour before the experiment.
1. Calculate the total volume of the working solution: 100µl / well × quantity of wells (Allow 100-200µl more than the total volume).
2. Dilute the biotin-detection antibody with antibody dilution buffer at 1:100 and mix thoroughly (i.e. Add 1µl of biotin-labelled antibody into 99µl of antibody dilution buffer.)
Figure 7. Preparation of Biotin working solution.
Preparation of biotin-labelled antibody working solution:
Prepare within 30 minutes of starting the experiment.
1. Calculate the total volume of the working solution: 100µl / well × quantity of wells. (Allow 100-200µl more than the total volume)
2. Dilute the SABC with SABC dilution buffer at 1:100 and mix thoroughly. (i.e. Add 1µl of SABC into 99µl of SABC dilution buffer.)
Figure 8. Preparation of SABC working solution.
Assay Procedure
Before adding to the wells, equilibrate the SABC working solution and TMB substrate for at least 30 mins at 37°C. When diluting samples and reagents, ensure they are mixed completely and evenly. It is recommended to plot a standard curve for each test.
Figure 9. Sandwich ELISA Protocol workflow.
| Assay Procedure Steps |
| 1. Set standard, test sample (diluted at least ½ with Sample Dilution Buffer) and control (zero) wells on the pre-coated plate and record their positions. It is recommended to measure each standard and sample in duplicate. Wash plate 2 times before adding standard, sample and control (blank) wells. |
| 2. Aliquot 100µl of standard solutions into the standard wells. |
| 3. Add 100µl of Sample / Standard dilution buffer into the control (blank) well. |
| 4. Add 100µl of properly diluted sample (serum, plasma, tissue homogenates and other biological fluids) into test sample wells. |
| 5. Seal the plate with a cover and incubate at 37 °C for 90 mins. |
| 6. Remove the cover, aspirate the liquid from the plate and wash plate 2 times with Wash |
| 7. Add 100µl of Biotin-lablelled antibody working solution to the bottom of each well (standard, test sample & zero wells) without touching the side walls. |
| 8. Seal the plate with a cover and incubate at 37°C for 60 mins. |
| 9. Remove the cover, and wash plate 3 times with Wash buffer. |
| 10. Add 100µl of SABC working solution into each well, cover the plate and incubate at 37°C for 30 mins. |
| 11. Remove the cover and wash plate 5 times with Wash buffer. |
| 12. Add 90 µl of TMB substrate into each well, cover the plate and incubate at 37°C in dark for 10-20 mins. (Note: This incubation time is for reference only, the optimal time should be determined by the end-user.) As soon as a blue colour develops in the first 3-4 wells (with most concentrated standards) and the other wells show no obvious colour, terminate the reaction by moving to Step 13. |
| 13. Add 50µl of Stop solution into each well and mix thoroughly. The colour changes into yellow immediately. |
| 14. Read the O.D. absorbance at 450 nm in a microplate reader immediately after adding the stop solution. |